Notification that instrument is not
level�
Cause: The instrument is out of level and needs to be re-leveled.
Resolution: Call Technical Support for instructions on how to re-level the VALENT.
Slide tray cancellations�
Cause: Trough not filling for AR step.
Resolution: Check carboy tubing connections. Use the Drain Waste & Prime buffers tool
to confirm proper connection (even flow of solution with minimal bubbles in the line). If
all tubing is connected properly and solution is not flowing as expected, call Technical
support.
Cause: Temperature did not reach set point.
Resolution: Call Technical Support.