11
Lithium Battery
This device contains one or more Lithium Iron Magnesium Phosphate Batteries. The batteries should not be opened or burned. The
batteries are safe with normal use. Exposure to the ingredients contained within and or their combustion products could be harmful.
Risk of exposure occurs only if the battery is mechanically, thermally, or electrically abused to the point of compromising the battery
enclosure. If the battery enclosure is compromised, exposure to the electrolyte solution contained within can occur by inhalation,
ingestion, eye contact and skin contact. More information is available from the manufacture of the battery, LithiumWerks.
Safely Turning on the Cart
If the VALENT Cart must has to be switched on or off, the instrument must be switched on or off in a specific order to avoid system
failure. Instances where the instrument cart may need to be turned on or off include:
• When moving the instrument to a different location.
• If there has been a power outage and the instrument requires proper shut down.
• When the user does not plan to use the instrument for at least two weeks or longer a week or more.
• If the VALENT Software and/or printers are not functioning appropriately.
Turn ON the Cart
TO INSTRUMENT
CART BATTERY
Turn cart battery OFF
when cart power is off or
cart is unplugged.
CART POWER
COMPUTER
INSTRUMENT
POWER
Power: 1,800 VA
115 VAC 100/120 VAC
Model VLT8102: 50 Hz
Model VLT8103: 60 Hz
Circuit Breaker: ISA Resettable
WARNING
1. Turn On the Cart Power Circuit Breaker
2. Turn On the Cart Battery Circuit Breaker
3. Turn On the Instrument Power Switch Circuit Breaker
1
2
3
• Confirm that all circuit breakers on the side of the cart are in the OFF (down) position, (refer to the diagram above.)
• Confirm that the cart is plugged into a dedicated electrical outlet capable of supply at least 1850W continuously.
• Confirm that the 30A instrument power umbilical from the cart is connected to the receptacle on the right side of the instrument.
• Confirm that the battery sense cable from the cart is connected to the receptacle on the right side of the instrument.
• Turn on the cart power circuit breaker, which is the white switch located on the electrical panel on the right side of the cart.
• Wait 60 seconds for the battery charger to sense wall power.
• Turn on the cart battery circuit breaker, which is the black switch located on the right side of the panel.
• Immediately after the cart battery circuit breaker is turned on, the battery charger fan inside the cart will be audible for approximately
5 seconds, indicating the charger has turned on.
• If the charger does not turn on, toggle the cart battery circuit breaker off and then back on again until the charger turns on.
• Turn on the instrument power switch circuit breaker, which is the black switch located to left of on the left side of the panel.
• Turn on the instrument using the large white power switch on the right side of the instrument.
• Turn on the computer.