Operating Instructions – MNT
2020-03-020.00
19
Disinf. delay
10 h
Select temperature increase days and set the days at
that the automatic power saving mode will be active
for 24 hours.
(e.g. Saturdays and Sundays in this example)
Incr. start
21:00
Incr. end
04:00
>
Incr. days
_ _ _ _ _ 6 7
1-Monday 2-Tuesday 3-Wednesday 4-Thursday 5-Friday 6-Saturday 7-Sunday
Settings for disinfection parameters:
The disinfection cycle works in a semi-automated
way.
First select a time when the system will ask for a
disinfection cycle.
(e.g. at 12:00 h the instrument will ask for a
disinfection cycle).
>
Disinfec. Req.
12:00
Disin. Days
1 2 _ _ _ _
Disin. Delay
10 h
Incr. start
21:00
Disinfec. Req.
12:00
Select the days in which a disinfection cycle can be
activated for the whole day.
(e.g. Monday and Tuesday in this example)
>
Disin. Days
1 2 _ _ _ _
Disin. Delay
10 h
Incr. start
21:00
1-Monday 2-Tuesday 3-Wednesday 4-Thursday 5-Friday 6-Saturday 7-Sunday
Select the delay time at which the disinfection cycle
will be activated.
(e.g. 10 hours)
After confirmation the disinfection cycle will start at
10 p.m.
Disinfec. Req.
12:00
Disin. Days
1 2 _ _ _ _
>
Disin. Delay
10 h
Incr. start
21:00