Operating Instructions – MNT
2020-03-020.00
29
8.12 MOTOR SETTING
For setting of vertical motor speeds, press
to
activate the motor.
TUE 13:05:05
CUT: 7 µm
Trim:
Cha: -20 °C (-22)
QF: 0
Status:OK
0000
The display will indicate the speed for the downward
(
↓
) and the upward (
↑
) movement of the specimen.
TUE 13:05:05
CUT: 7 µm
Trim: 19µm
Cha: -20 °C (-22)
QF: 0
↓
40 mm/s
↑
125 mm/s
To adjust the setting for the downward movement,
press
and simultaneously
or
to increase or
decrease the speed setting.
TUE 13:05:05
CUT: 7 µm
Trim: 19µm
Cha: -20 °C (-22)
QF: 0
↓
35 mm/s
↑
125 mm/s
To adjust the setting for the upward movement, press
and simultaneously
or
to increase or
decrease the speed setting.
TUE 13:05:05
CUT: 7 µm
Trim: 19µm
Cha: -20 °C (-22)
QF: 0
↓
40 mm/s
↑
140 mm/s