Operating Instructions – MNT
2020-03-020.00
11
Freezing bar
Minimum temperature
10 K lower than chamber temperature
Number of positions
24
(Standard version)
resp. 21 + 2
(version with quick freeze
places)
Quick freeze positions
2 (optional), to -55 °C (Peltier cooling)
Microtome
Type
rotary microtome, incapsulated, motorized
advance, motorized cutting
Thickness range
0.5 – 100 µm
Thickness settings
0 – 2 µm in 0.5 µm steps
2 – 20 µm in 1 µm steps
20 – 50 µm in 2 µm steps
50 – 100 µm in 5 µm steps
Horizontal movement
28 mm
Vertical movement
58 mm
Retraction during upward stroke
0 – 200 µm, free selectable
(defined steps)
Trimming
0,5 – 750 µm, free selectable
(defined steps)
Specimen orientation
8° (x and y axes), 360° (z axes)
Zero point detent x and y axis
Motor speed
3 to 300 mm/s
Cryostat
Dimensions (L x W x H)
680 mm x 760 mm x 1110 mm
Weight (without accessories)
135 kg (without object cooling)
154 kg (with object cooling)
*All temperature specifications refer to an ambient temperature of + 20 ° C and a relative humidity of 60%