KBF + KMF (E6) 07/2017
page 42/155
5.8
Performance when opening the door
When you open the door the fan starts running with minimum speed.
After 60 seconds from opening the door, heating, refrigeration, humidification, dehumidification and fan
turn off.
After closing the door, heating, refrigeration, humidification, dehumidification and fan turn on again.
6. Start up
6.1
Turning on the chamber
After connecting the supply lines (chap. 4), turn on the chamber by its main power switch (1). The lit pilot
lamp shows the chamber is ready for operation.
When the main power switch is turned on and yet the controller display is dark, the display is in stand-by
mode. Press on the touchscreen to activate it.
•
Open the water-tap for freshwater supply. Alternatively, fill the freshwater can (option, chap. 19.9).
•
The humidifying and dehumidifying system must be activated (deactivated operation line “Humidity
off”, chap. 7.3 and setting “Control on”, chap. 6.3).
After the first turning on of the chamber or after an interruption of the power supply the relative humidity
will increase after a delay of about 20 minutes. During this period, the relative humidity can drop consid-
erably.
Warming chambers may release odors in the first few days after commissioning. This is not a quality
defect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for
one day and in a well-ventilated location.
WARNING: If customer should use a BINDER chamber running in non-supervised continu-
ous operation, we strongly recommend in case of inclusion of irrecoverable specimen or
samples to split such specimen or samples and store them in at least two chambers, if this is
feasible.
6.2
Controller settings upon start up
The window „Language selection“ enables the
language selection
, in case that it’s activated in the
“Start-up” menu. Afterwards occurs a request of the
time zone
and the
temperature unit
.
The controller will function in the
operating mode
, which was active before the last shut-down. It controls
temperature and humidity in fixed value operating mode to the last entered values and in the program
mode to the set points achieved beforehand.
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...