KBF + KMF (E6) 07/2017
page 25/155
4.3
Freshwater supply
Connect the waste water pipe before connecting the chamber to a freshwater pipe or filling the
freshwater can (option, chap. 19.9).
You can supply the chamber with freshwater via a water pipe or by manually filling a freshwater can (op-
tion, chap. 19.9).
Water intake temperature NOT below +5 °C /
41 °F
and not exceeding 40 °C /
104 °F
.
CAUTION
Calcification of the humidifying system.
Damage to the chamber.
Operate the chamber with deionized (demineralized) water only.
Types of suitable water quality:
•
Deionized water from a water treatment installation already existing at the customer's site. Conductivi-
ty from 1 µS /cm up to a maximum of 20 µS/cm. (Water, which is in equilibrium with the CO
2
in the air,
and has a conductivity below 1 µS/cm (ultrapure water), may cause acid corrosion due to its low pH.)
•
Water treated by the optional water treatment system BINDER Pure Aqua Service (disposable sys-
tem). A reusable measuring equipment to assess the water quality is included (chap. 19.10).
BINDER GmbH is NOT responsible for the water quality at the user’s site.
Any problems and malfunctions that might arise following use of water of deviating quality is
excluded from liability by BINDER GmbH.
The warranty becomes void in the event of use of water of deviating quality.
4.3.1 Automatic fresh water supply via water pipe
An enclosure inside the chamber contains the connection kit for freshwater and wastewater. Install the
freshwater connection using either the enclosed water hose or another pressure-resistant one. To ac-
complish this, remove the cover of the freshwater connection “IN” (13) on the rear of the chamber. Pro-
tect both ends of the hose with two of the four supplied hose clamps.
Before turning on the chamber, check the connection for leaks. Water supply is automatically effected via
the freshwater connection “IN” (13).
As the appliance only lets in water when required, there is no continuous water flow.
•
Supply pressure 1 to 10 bar when connecting to a water pipe
•
Water type: deionized (demineralized) water
•
Water intake temperature NOT below +5 °C /
41 °F
and not exceeding 40 °C /
104 °F
.
•
The water intake should be provided with a shut-off slide or water-tap.
•
For the water supply, fix the delivered adapter with hose olive on the thread at the rear of
the chamber.
•
Protect the water supply at one side with the supplied hose clamp.
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...