KBF + KMF (E6) 07/2017
page 29/155
4.4
Electrical connection
The chambers are supplied ready for connection. They come with a fixed power connection cable of at
least 1800 mm /
70.87 in
in length.
Model version
Art. No.
(x = 0 or 1)
Power plug
V/-10% at the
indicated power fre-
quency
Current
type
Chamber
fuse
KBF115-230V
KBF240-230V
KBF720-230V
KBF1020-230V
9x20-0320
9x20-0322
9x20-0324
9x20-0326
Shock-proof plug
200-230 V at 50 Hz
200-230 V at 60 Hz
1N~
16 Amp
KBF115-240V
KBF240-240V
KBF720-240V
KBF 1020-240V
9x20-0321
9x20-0323
9x20-0325
9x20-0327
NEMA 6-20P
200-240 V at 50Hz
200-240 V at 60Hz
2~
16 Amp
KMF115-230V
KMF240-230V
KMF720-230V
9x20-0341
9x20-0343
9x20-0345
Shock-proof plug
200-230 V at 50 Hz
200-230 V at 60 Hz
1N~
16 Amp
KMF115-240V
KMF240-240V
KMF720-240V
9x20-0342
9x20-0344
9x20-0346
NEMA 6-20P
200-240 V at 50Hz
200-240 V at 60Hz
2~
16 Amp
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the lat-
est technology. The protective conductors of the socket and plug must be compatible!
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (left chamber side, bottom right-hand, see chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company
and as well as the VDE directives (for Germany). We recommend the use of a residual current circuit
breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Installation category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 23.5).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Remark when operating the chamber with a power frequency of 60 Hz:
WARNING
High leakage current.
Electrical hazard.
Earth connection is essential before connecting supply. Check socket before inserting
plug.
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...