KBF + KMF (E6) 07/2017
page 132/155
Chamber size
115
240
720
1020
Racks
Quantity of racks (regular)
2
2
2
2
Quantity of racks (max.)
5
9
15
15
Maximum load per rack
kg /
lbs.
30 /
66
30 /
66
45 /
99
45 /
99
Maximum permitted total load
kg /
lbs.
100 /
220
100 /
220
150 /
331
150 /
331
Weight
Weight (empty)
kg /
lbs.
129 /
284
184 /
406
309 /
681
365 /
805
Temperature data (without humidity)
Temperature range
°C /
°F
0 to +70 /
32 to 158
0 to +70 /
32 to 158
0 to +70 /
32 to 158
0 to +70 /
32 to 158
Temperature fluctuation
± K
0.1
0.1
0.1
0,5
Temperature
uniformity (variation)
at 25 °C /
77 °F
± K
0.2
0.2
0,2
0,2
at 40 °C /
104 °F
± K
0.2
0.3
0,2
0,2
Max. heat compensation at 40 °C /
104 °F
W
200
300
600
600
Climatic data (with humidity)
Temperature range
°C / °F
+10 to +70 /
50 to 158
+10 to +70 /
50 to 158
+10 to +70 /
50 to 158
+10 to +70 /
50 to 158
Temperature
fluctuation
at 25 °C /
77 °F
/ 60% r.H.
± K
0.1
0.1
0.1
0,1
at 40 °C /
104 °F
/ 75% r.H.
± K
0.1
0.1
0.1
0,1
Temperature
uniformity
(variation)
at 25 °C /
77 °F
/ 60% r.H.
± K
0.2
0.3
0.2
0,2
at 40 °C /
104 °F
/ 75% r.H.
± K
0.2
0.3
0.2
0,2
Humidity range
% r.H.
10 to 80
10 to 80
10 to 80
10 to 80
Humidity
fluctuation
at 25 °C /
77 °F
/ 60% r.H. ± % r.H.
≤ 2
1.5
1.5
≤ 1,5
at 40 °C /
104 °F
/ 75% r.H. ± % r.H.
≤ 2
1.5
1.5
≤ 1,5
Recovery
time after
doors were
open for 30 s
at 25 °C /
77 °F
/ 60% r.H. minutes
6
5
16
11
at 40 °C /
104 °F
/ 75% r.H. minutes
7
11
17
12
Electrical data
(model versions KBF115-230V, KBF240-230V, KBF720-230V, KBF1020-230V)
IP system of protection acc. to EN 60529
20
20
20
20
Nominal volt-
age (+/-10%)
at 50 Hz power frequency
V
200-230
200-230
200-230
200-230
at 60 Hz power frequency
V
200-230
200-230
200-230
200-230
Current type
1N~
1N~
1N~
1N~
Power plug
shock proof plug
Nominal power
kW
2.00
2.10
3.10
3.10
Installation category acc. to IEC 61010-1
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
Over-current release category B, 2 poles
Amp
16
16
16
16
Different electrical data for KBF constructed for the USA and Canada
(model versions KBF115-240V, KBF240-240V, KBF720-240V, KBF1020-240V)
Nominal volt-
age (+/-10%)
at 50 Hz power frequency
V
200-240
200-240
200-240
200-240
at 60 Hz power frequency
V
200-240
200-240
200-240
200-240
Current type
2~
2~
2~
2~
Power plug
NEMA
6-20P
6-20P
6-20P
6-20P
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...