KBF + KMF (E6) 07/2017
page 134/155
Chamber size
115
240
720
Temperature data (without humidity)
Temperature range
°C / °F
-10 to +100
/
14 to 212
-10 to +100
/
14 to 212
-10 to +100
/
14 to 212
Average heating up time acc. to IEC 60068-3-5
K/min.
1.3
0.8
0.7
Average cooling down time acc. to IEC 60068-3-5
K/min.
0.5
0.4
0.4
Heating up time
from -10 °C /
14 °F
to +100 °C /
212 °F
minutes
85
140
155
Cooling down time
from +100 °C /
212 °F
to -10 °C /
14 °F
minutes
240
360
350
Max. heat compensation at 25 °C /
77 °F
W
150
200
450
Climatic data (with humidity)
Temperature range
°C / °F
+10 to +90 /
50 to 194
+10 to +90 /
50 to 194
+10 to +90 /
50 to 194
Temperature fluctuation *)
± K
0.1 to 0.3
0.1 to 0.3
0.1 to 0.5
Temperature uniformity (variation) *)
± K
0.2 to 1.0
0.1 to 1.0
0.1 to 1.0
Humidity range
% r.H.
10 to 98
10 to 98
10 to 98
Humidity fluctuation *)
± % r.H.
≤ 2,5
≤ 2
≤ 2
Dew point temperature range
°C
+5 to +90
+5 to +90
+5 to +90
Electrical data
(model versions KMF115-230V, KMF240-230V, KMF720-230V)
IP system of protection acc. to EN 60529
20
20
20
Nominal voltage
(+/-10%)
at 50 Hz power frequency
V
200-230
200-230
200-230
at 60 Hz power frequency
V
200-230
200-230
200-230
Current type
1N~
1N~
1N~
Power plug
Shock-proof plug
Nominal power
kW
2.00
2.10
3.10
Installation category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Over-current release category B, 2 poles
Amp
16
16
16
Deviant electrical data KMF for the USA and Canada
(model versions KMF115-240V, KMF240-240V, KMF720-240V)
Nominal voltage (+/-10%)
at 50 Hz power frequency
V
200-240
200-240
200-240
at 60 Hz power frequency
V
200-240
200-240
200-240
Current type
2~
2~
2~
Power plug
NEMA
6-20P
6-20P
6-20P
Environment-specific data
Noise level (mean value)
dB (A)
52
52
56
Energy consumption at 85 °C /
185 °F
and 85 % r.H.
Wh/h
570
570
900
Filling weight of refrigerant R 134A (GWP 1300)
kg
0.180
0.240
0.430
*)
Depending on the set-point.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3°C /
71.6 °F +/- 5.4 °F
and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER factory standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right
to change technical specifications at any time.
Refrigerating performance decreases while operating the chamber at temperatures < 0 °C /
32 °F
due to icing of the evaporators. For this reason defrost the chamber regularly, e.g. once
a week.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...