3.2
Guidelines for safe lifting and transportation
The front castors of KMF 240, 720 and 1020 can be blocked by brakes. After operation, please observe
the guidelines for temporarily decommissioning the chamber (chap. 21.2). Please move the chambers
with castors only when empty and on an even surface, otherwise the castors may be damaged.
CAUTION
Sliding or tilting of the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
Transport the chamber in its original packaging only.
For moving or shipping, secure the constant climate chamber with transport straps.
∅
Do NOT lift or transport the chamber using the door, the handle, or the lower housing.
Lift chambers size 115 at the four lower corners with the aid of 4 people
Lift chambers size 240 at the four lower corners with the aid of 6 people or with a fork
lifter. Set the fork lifter only from the front or rear in the middle of the chamber.
Lift the chambers sizes 720 and 1020 using technical devices (fork lifter). Set the fork
lifter only from the front or rear in the middle of the chamber.
∅
Do NOT set the fork lifter from the chamber side.
You can order transport packing for moving or shipping purposes from BINDER service.
Permissible ambient temperature range during transport:
•
If the steam humidifying system has NOT been emptied: +3 °C /
37.4 °F
to +60 °C /
140 °F
.
•
After BINDER Service has emptied the steam humidifying system: -10 °C /
14 °F
to +60 °C /
140 °F
.
With temperatures below +3 °C /
37.4 °F
, water must be completely removed from the humidifying sys-
tem.
CAUTION
Transport below +3 °C / 37.4 °F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the chamber.
Contact BINDER Service before any transportation below +3 °C /
37.4 °F
.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 21.2).
Permissible ambient temperature range during storage:
•
If the steam humidifying system has NOT been emptied: +3 °C /
37.4 °F
to +60 °C /
140 °F
.
•
After BINDER Service has emptied the steam humidifying system: -10 °C /
14 °F
to +60 °C /
140 °F
Summary of Contents for KBF 115
Page 137: ...KBF KMF E6 07 2017 page 137 155 23 8 Dimensions size 115 ...
Page 138: ...KBF KMF E6 07 2017 page 138 155 23 9 Dimensions size 240 ...
Page 139: ...KBF KMF E6 07 2017 page 139 155 23 10 Dimensions size 720 ...
Page 140: ...KBF KMF E6 07 2017 page 140 155 23 11 Dimensions size 1020 ...
Page 142: ...KBF KMF E6 07 2017 page 142 155 ...
Page 143: ...KBF KMF E6 07 2017 page 143 155 ...
Page 144: ...KBF KMF E6 07 2017 page 144 155 24 2 EU Declaration of Conformity for KMF ...
Page 145: ...KBF KMF E6 07 2017 page 145 155 ...
Page 146: ...KBF KMF E6 07 2017 page 146 155 ...
Page 148: ...KBF KMF E6 07 2017 page 148 155 ...
Page 149: ...KBF KMF E6 07 2017 page 149 155 25 Product registration ...