KB (E4 + E6) 02/2017
page 21/148
3.4
Location of installation and ambient conditions
Set up the chamber on a flat, even surface, free from vibration, in a well-ventilated, dry location and align
it using a spirit level. The site of installation must be capable of supporting the chamber’s weight (see
technical data, chap. 23.4). The chambers are designed for setting up inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the chamber.
∅
Do NOT set up chambers in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation: +18 °C /
64.4 °F
to +32 °C /
89.6 °F
. At ele-
vated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +22 °C +/- 3 °C /
71.6 °F ± 5.4 °F
to which the specified technical data relates.
Deviations from the indicated data are possible for other ambient conditions. Lower values of
the temperature range indicated in the technical data are valid at an ambient temperature of
max. 25 °C
/ 77 °F
.
With each degree of ambient temperature > +25 °C /
77°F
, the refrigeration power decreases
by 1.5 K.
•
Permissible ambient humidity: 70 % r.H. max., non-condensing.
When operating the chamber at temperature set-points below ambient temperature, high ambient humidi-
ty may lead to condensation on the chamber.
•
Installation height: max. 2000 m /
6562 ft.
above sea level.
When placing several chambers of the same size side by side, maintain a minimum distance of 250 mm /
9.84 in
between each chamber. Wall distances: rear 100 mm /
3.94 in
, sides 160 mm / 6
.29 in
. Spacing
above the chamber of at least 100 mm /
3.94 in
must also be maintained.
Two KB 53 / 115 (E4) chambers of the same size can be piled on top of each other. For this purpose
place rubber pads under every foot of the upper chamber to prevent the device from slipping.
CAUTION
Sliding or tilting of the upper chamber.
Damage to the chambers.
When stacking, place rubber pads under every foot of the upper chamber.
Stack only chambers of the same size..
KB 240 / 400 / 720 (E6) chambers must NOT be stacked.
CAUTION
Danger by stacking.
Damage to the chambers.
∅
Do NOT place the cooling incubators on top of each other.
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user
there is no risk
of
temporary
overvoltages
in the sense of EN 61010-1:2010.
Summary of Contents for KB 115 (E4)
Page 133: ...KB E4 E6 02 2017 page 133 148 23 7 KB 53 dimensions mm...
Page 134: ...KB E4 E6 02 2017 page 134 148 23 8 KB 115 dimensions...
Page 135: ...KB E4 E6 02 2017 page 135 148 23 9 KB 240 dimensions mm...
Page 136: ...KB E4 E6 02 2017 page 136 148 23 10 KB 400 dimensions mm...
Page 137: ...KB E4 E6 02 2017 page 137 148 23 11 KB 720 dimensions...
Page 139: ...KB E4 E6 02 2017 page 139 148...
Page 141: ...KB E4 E6 02 2017 page 141 148...
Page 142: ...KB E4 E6 02 2017 page 142 148 25 Product registration...
Page 149: ......