20. Maintenance, cleaning, and service
20.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of death.
∅
The chamber must NOT become wet during operation or maintenance works.
∅
Do NOT remove the rear panel of the chamber.
Before conducting maintenance work, turn off the chamber at the main power switch
and disconnect the power plug.
General maintenance work must be conducted by licensed electricians or experts au-
thorized by BINDER.
Maintenance work at the refrigeration system must only be conducted by qualified per-
sonnel who underwent training in accordance with EN 13313:2010 (e.g. a refrigeration
technician with certified expert knowledge acc. to regulation 303/2008/EC). Follow the
national statutory regulations.
Ensure regular maintenance work is performed at least once a year and that the legal requirements are
met regarding the qualifications of service personnel, scope of testing and documentation. All work on the
refrigeration system (repairs, inspections) must be documented.
The warranty becomes void if maintenance work is conducted by non-authorized personnel.
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged.
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) sev-
eral times a year.
We recommend taking out a maintenance agreement. Please consult BINDER Service.
BINDER telephone hotline:
+49 (0) 7462 2005 555
BINDER fax hotline:
+49 (0) 7462 2005 93555
BINDER e-mail hotline:
BINDER service hotline USA:
+1 866 885 9794 or +1 631 224 4340 x3 (toll-free in the USA)
BINDER service hotline Asia Pacific:
+852 390 705 04 or +852 390 705 03
BINDER service hotline Russia and CIS
+7 495 988 15 16
BINDER Internet website
http://www.binder-world.com
BINDER address
BINDER GmbH, post office box 102, D-78502 Tuttlingen
International customers, please contact your local BINDER distributor.
Summary of Contents for KB 115 (E4)
Page 133: ...KB E4 E6 02 2017 page 133 148 23 7 KB 53 dimensions mm...
Page 134: ...KB E4 E6 02 2017 page 134 148 23 8 KB 115 dimensions...
Page 135: ...KB E4 E6 02 2017 page 135 148 23 9 KB 240 dimensions mm...
Page 136: ...KB E4 E6 02 2017 page 136 148 23 10 KB 400 dimensions mm...
Page 137: ...KB E4 E6 02 2017 page 137 148 23 11 KB 720 dimensions...
Page 139: ...KB E4 E6 02 2017 page 139 148...
Page 141: ...KB E4 E6 02 2017 page 141 148...
Page 142: ...KB E4 E6 02 2017 page 142 148 25 Product registration...
Page 149: ......