Function overview of screen program controller MB1
Pilot lamp: Ready for operation
EXIT button (to exit a menu point)
AUTOMATIC button (to start a previously entered program)
ENTER button (to confirm a selection)
Navigation buttons (functions are assigned by the menu)
Figure 6: Display program controller MB1
The program controller MB1 controls the temperature inside the high performance temperature chamber
(range: 5 °C above ambient temperature up to 300 °C).
You can enter the desired set point values in Manual Mode or Program Mode (chap. 5.2) in the display
controller.
W
X
08:43:55 15.12.05
CONFIG
VIEW->
PGM
TEMP
40.0
36.8
°C
HAND
Set point value Actual value
Figure 7: Normal display of the MB1 program controller in Manual mode
5.2
Operating modes
The program controller MB1 operates in 3 modes:
Idle Mode
The controller is not functional, i.e., there is no heating. The fan turns
at 50% rate.
Manual Mode
(Fixed value operation)
(HAND)
The controller operates as a fixed-point control, i.e., a temperature
set-point can be defined, which is then maintained (chap. 8).
Program Mode (AUTO)
An entered temperature program is run (chap. 9).
The program controller MB1 allows programming temperature cycles.
The controller offers 25 program memory positions with 100 program sections each. The total number of
program sections of all programs is limited to 500.
Programming can be done directly through the keypad of the controller or graphically through the soft-
ware APT-COM™ 3 DataControlSystem (option, chap. 11.1) specially developed by BINDER.
Summary of Contents for APT.line M Series
Page 54: ...M 01 2009 page 54 61 15 6 Dimensions M 53 ENT ENT ENT...
Page 55: ...M 01 2009 page 55 61 15 7 Dimensions M 115 ENT ENT ENT...
Page 56: ...M 01 2009 page 56 61 15 8 Dimensions M 240 6 8 ENT ENT ENT...
Page 57: ...M 01 2009 page 57 61 15 9 Dimensions M 400 ENT ENT ENT...
Page 58: ...M 01 2009 page 58 61 15 10 Dimensions M 720 ENT ENT ENT...