M 01/2009
page 31/61
Program entry as set-point ramp (example)
On
Off
t/min.
Operation line 1 = air flap
W/°C
t/min.
01 02
03
04
0
50
100
150
200
250
30 120
180
380
05
Tolerance
minimum
Tolerance
maximum
Program table corresponding to the diagram above:
Program
section
Set-point
temp.
Fan Section
time
Operation
line1
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01 50
100
%
00:30:00
Off 1
0
-1999
+9999
02 100
100
%
01:30:00
Off 1 0 -5 +5
03 200
100
%
01:00:00
Off 1 0 -2 +2
04 200
100
%
03:20:00
Off 1 0
-1999
+9999
05 100
100
%
00:00:01
Off 1 0
-1999
+9999
Enter now the values of the above program table into the program table of one of the 25 program places
of the program controller MB1:
PGM
08:43:55 15.02.02
PGM
08:43:55 15.12.05