M 01/2009
page 29/61
Hit the PGM button. An inquiry screen appears offering to enter or delete individual program sections:
ZP-Abschnitt Abs. Nr. 5
insert
delete
In this view, new program lines can be entered or deleted:
new
New lines are added below in the table
insert
New lines are added above a previously selected line
delete
Individual lines that have been selected previously are deleted
Create as many lines, i.e. program sections, as desired. In a next step, values can be entered into these
lines. It is as well possible to add supplementary lines later on or to delete individual lines at any moment.
PGM
08:43:55 15.02.02
PGM
08:43:55 15.12.05
To enter values, select the corresponding line via the
arrow keys.
Hit the ENTER button. The
program editor
appears.
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****.*
Setpoint 1
+100.0
FAN
Operating contacts
Time
00:45:00
Repeat Section
5
Repeat Number
10
Tol.-band min.
-1999.0
Tol.-band max.
+9999.0
Parameter set
1
-- Temperature at the start of the program section
-- Fan speed in %
-- Operating
contact
(operation
line)
1
=
air flap open / closed
-- Duration of the program section
-- No. of start section in case of repeat cycles
-- No. of repeats in case of repeat cycles
-- Temperature limits (maximum / minimum temperature)
In case of exceeding: temporary program stop.
-- Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by ENTER.
Then enter the values via the arrow keys, and confirm the entry by ENTER.
Adapt the temperature safety device class 2 (chap. 10.1) or the temperature safety device
class 3.1 (option, chap. 10.2) to the highest temperature set-point value of the program actu-
ally used. Check the safety device for each temperature program and adapt it if necessary.
Set the set-point of temperature safety device class 2 or class 3.1 (option) by about 5 °C to 10
°C above the controller temperature set-point.
Behavior after completing the program:
The controller changes to Idle Mode. The heating is inactive; the chamber approximates ambient tem-
perature. The fan turns at 50% rate.
Summary of Contents for APT.line M Series
Page 54: ...M 01 2009 page 54 61 15 6 Dimensions M 53 ENT ENT ENT...
Page 55: ...M 01 2009 page 55 61 15 7 Dimensions M 115 ENT ENT ENT...
Page 56: ...M 01 2009 page 56 61 15 8 Dimensions M 240 6 8 ENT ENT ENT...
Page 57: ...M 01 2009 page 57 61 15 9 Dimensions M 400 ENT ENT ENT...
Page 58: ...M 01 2009 page 58 61 15 10 Dimensions M 720 ENT ENT ENT...