M 01/2009
page 27/61
Setting ranges:
Temperature
0 °C up to 300 °C
Fan speed
0 % to 100 %
Fan speed can be reduced to standstill of the fan. Do this only if needed, because the
spatial distribution of temperature is as well reduced.
Technical data refer to 100% fan
speed.
Adapt the temperature safety device class 2 (chap. 10.1) or the temperature safety device
class 3.1 (option, chap. 10.2) every time the set-point for temperature is changed.
Set the set-point of temperature safety device class 2 or class 3.1 (option) by about 5 °C to 10
°C above the controller temperature set-point.
If operation line 1 has been set to ON, i.e., the air flap is open,
the notification “AIR FLAP OPEN” is displayed on the controller
MB1 display next to a flashing blue information symbol.
In Manual Mode, no program can be started. A set-point can be entered for temperature. The actual val-
ue equilibrates to this set-point.
When pushing the EXIT button in Manual Mode, the controller changes to Idle Mode. The set-points en-
tered in Manual Mode remain saved.
When incidentally pressing the EXIT or AUTOMATIC button during Manual Mode operation,
the controller will change to Idle Mode and thus will not adjust any more to the program set-
points.
We recommend keyboard locking (option, chap. 11.3.) during operation.
8.2
Behavior after power failure in Manual Mode
In Manual Mode (HAND), all functions reinstate exactly to the same situation the chamber had before
power failure. The set-point is immediately resumed. The power failure is noted in the event list. No error
message indicating that a power failure has taken place is displayed.
9.
Program operation
The 1-channel program controller MB1 allows programming temperature cycles. It offers 25 program
memory positions with 100 program sections each. The total of program sections of all programs is lim-
ited to 500. It is not possible to link several programs.
For each program section a temperature set-point, the fan speed (0% up to 100%), and the switching-
state of up to 8 operation lines can be entered. Operating line 1 is used to control the air flap position.
The other operation lines are non-functional.
Programming is possible directly by the keypad of the controller or graphically by the software APT-
COM™ 3 DataControlSystem (option, chap. 11.1) specially developed by BINDER.
HAND
W
VIEW->
AIRFLAP OPEN
Summary of Contents for APT.line M Series
Page 54: ...M 01 2009 page 54 61 15 6 Dimensions M 53 ENT ENT ENT...
Page 55: ...M 01 2009 page 55 61 15 7 Dimensions M 115 ENT ENT ENT...
Page 56: ...M 01 2009 page 56 61 15 8 Dimensions M 240 6 8 ENT ENT ENT...
Page 57: ...M 01 2009 page 57 61 15 9 Dimensions M 400 ENT ENT ENT...
Page 58: ...M 01 2009 page 58 61 15 10 Dimensions M 720 ENT ENT ENT...