FDL (E2) 03/2011
page 62/68
Temperature data
Temperature range, 5 °C above ambient up to
°C /
°F
300
/
572
Maximum temperature of the heating surface
°C /
°F
750
/
1382
Maximum permitted oven temperature
°C /
°F
300
/
572
Temperature fluctuation
± K
0.3
at 70 °C /
158°F
± K
1.5
at 150 °C /
302°F
± K
2.5
Temperature uniformity (variation)
at 300 °C /
572°F
± K
4
to 70 °C /
158°F
min 7
to 150 °C /
302°F
min 17
Heating-up time 2)
to 300 °C /
572°F
min 44
to 70 °C /
158°F
min 1
to 150 °C /
302°F
min 3
Recovery time after door was opened
for 30 sec. 2)
to 300 °C /
572°F
min 6
Ventilation data
Air change
approx. x/min.
3
Air circulation
approx. x/min.
40
approx. l/min
400
Minimum volumetric flow rate of exhaust air 3)
approx.
m
3
/h 24.0
Air flow rate
m/sec
0.8 up to 1.2
Solvent data
Highest permitted solvent quantity 4)
(at T-180 °C, M-100g/mol, U-40g/m3, K=0,5)
g 6.65
Electrical data
Housing protection acc. to EN 60529
IP
33
Nominal voltage (+10 %) 50/60 Hz
V
230 1 N ~
Nominal current
Amp
13.0
Nominal power
kW
2.90
Sound-pressure level
dB(A)
< 70
at 70 °C /
158°F
Wh/h 388
at 150 °C /
302°F
Wh/h 1098
Energy consumption 5)
at 300 °C /
572°F
Wh/h 2023
Power plug
shock proof plug
Over-voltage category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Legend:
1) If the volume of the dry goods inserted in the dryer exceeds 10% of the total steam area, it must be
deducted in the calculation of the total steam volume.
2) Up to 98 % of the set value
3) Procedure: "calculation of the volumetric flow rate through the exhaust duct based on the velocity of
flow measured there with a propeller anemometer "
4) At a drying temperature of 180 degrees Celsius and an average molecular weight of the solvent of
M=100g/mole. For a different drying temperature, the highest permissible solvent quantity must be re-
calculated. The calculation should be made on the basis of the "principles for the calculation of ventila-
tion of chamber dryers and continuous dryers" in accordance with industrial standard EN 1539:2009,
Appendix B.
5) Use this value for sizing air condition systems.
All technical data is specified for units with standard equipment at an ambient temperature of +25 °C /
77°F
and a power supply voltage fluctuation of ±10. The temperature data is determined in accordance to
BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of
the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the cabinet is fully loaded, the specified heating up times may vary according to the load.