FDL (E2) 03/2011
page 54/68
14.2 Replacing the intake filter
The fresh air intake filter (fine particle filter for particles 1µm up to 10 µm, class F6/EU6 acc. to EN
779:2002) at the top right-hand side must be replaced from time to time, depending on the degree of
soiling. Pull out the slide and replace the filter cartridge.
Intake filter
Slide
Figure 15: Fresh air intake filter
14.3 Cleaning and decontaminating the safety / paint drying oven
DANGER
Electrical hazard.
Danger of death.
∅
Do NOT spill water or cleaning agents over the inner and outer surfaces.
Disconnect the chamber before cleaning. Disconnect the power plug.
Completely dry the appliance before turning it on again.
Cleaning
Always keep clean the inner parts of the safety / paint drying oven including the drip pans, drip
trays and exhaust-air ducts. Remove the remnants of coating material at suitable intervals.
Disconnect the chamber from the power supply before cleaning. Disconnect the power plug.
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces
inner chamber
racks
door gaskets
Standard commercial cleaning detergents free from acid or halogenides.
Alcohol based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Instrument panel
Standard commercial cleaning detergents free from acid or halogenides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
For surface protection, perform cleaning as quickly as possible.
After cleaning completely remove any cleaning agents from the surfaces with a moistened
towel.
Soapsuds may contain chlorides and must therefore NOT be used for cleaning.
Summary of Contents for 9010-0269
Page 6: ...FDL E2 03 2011 page 6 68...
Page 69: ......