Instructions for use - BW 585
BIEGLER
2 DESCRIPTION
2.1 GENERAL DESCRIPTION
The BIEGLER BW 585 is a warmer for infusions or transfusions and operates on
the basis of continuous flow heating, where the heat from the heat exchanger is
transferred via the extension tubing to the liquid flowing within it.
The patented groove design enables several extension tubes to be used,
provided that there is sufficient heating provided for each of the tubes placed in
position.
The design of the casing allows rapid and simple fitting to any suitable infusion
stand. The temperature of the heating element can be set between 37°C and
41°C in steps of 0.5°C and is displayed as an illuminated band.
The preset temperature after switching on the BW 585 is 38.5°C.
The alarm and self-test functions for high and low temperature that are
incorporated into the device assure safe operation.
2.2 SCOPE OF DELIVERY
Blood and infusion warmer BW 585 and instructions for use.
2.3 CONSUMABLE MATERIAL
Various consumable materials are available according to requirements.
Order numbers:
Extension set
3.5 m
FP1002001
Order no. 35000
Extension set
4.6 m
FP1760110
Order no. 45000
Extension set with bubble trap
4.6 m
FP4600001
Order no. 25000
4
EN - Edition 03 / 2007
BIEGLER
Instructions for use - BW 585
Table 202 – Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS (see 6.8.3.201 a) 6))
Guidance and manufacturer´s declaration – electromagnetic immunity
The BW 585 is intended for use in the electromagnetic environment specified below. The customer
or the user of the BW 585 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity
should be at least 30 %.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
not applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(>95 % dip in U
T
)
for 5 sec
< 5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the BW 585 requires continued
operation during power mains
interruptions, it is recommended
that the BW 585 be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE U
T
is the a. c. mains voltage prior to application of the test level.
EN - Edition 03 / 2007
13