Instructions for use - BW 585

BIEGLER

6 PERIODIC INSPECTIONS

Periodic technical safety inspections must be performed on the BW 585 at least
every 12 months by persons who on the basis of their training, knowledge and
practical experience are qualified to carry out such technical safety testing.

The results of the periodic inspection are to be entered, together with the date
and the inspecting agency, on the reverse side of the instructions for use. 

Important: If a malfunction is identified in the periodic inspection, affix a suitable
warning notice on the device to ensure that it can no longer be used until the
necessary repairs have been carried out.

CHECKING THE WARM-UP PERIOD

This is the time taken by the BW 585 to heat up to 41.0°C from room
temperature. The device is malfunctioning if it takes much longer than one
minute.

CHECKING THE CONTROL TEMPERATURE

The control temperature is checked on the groove bed of the heat exchanger.
The sensor of a suitable contact thermometer (tol/- 0.15°C) is fixed to
this place e.g. using a piece of infusion tubing. The check is performed at a
temperature setting of 38.5°C. The measured value is read after it has
stabilized. The difference must not /- 0.5°C. There is a malfunction if a
difference from control temperature of greater than +/- 0.5°C is obtained.

CHECKING THE LOW TEMPERATURE ALARM

Preheat the device to 38.5°C, then disconnect the mains plug. Hold down the
control 

 and reconnect the mains plug. Push the 

 switch. The device is now

in an operational state where all the alarms are active, but the heating is
switched off. The BW 585 now slowly cools down. When the temperature drops
below the 36.5°C threshold, the low temperature alarm should be active.
For reasons of safety, short beeping sounds are given at intervals of a second in
this operational mode and the LED ON and STANDBY indicators flash
alternately (Diag.1/4 and Diag.1/5). There is a malfunction if the low temperature
alarm is not triggered.

8

EN - Edition 03 / 2007

BIEGLER

Instructions for use - BW 585

CHECKING THE HIGH TEMPERATURE ALARM

Preheat the device to 41°C and wait for the temperature to stabilize, then
disconnect the mains plug. Hold down the control 

 and reconnect the mains

power plug. Push the 

 switch. The device now heats up to a target

temperature of 42.5°C. Observe the temperature indicator carefully. The high
temperature alarm should be triggered at a temperature of 42°C.
For reasons of safety, short beeping sounds are given at intervals of a second in
this operational mode and the LED ON and STANDBY indicators flash
alternately (Diag.1/4 and Diag.1/5). There is a malfunction if the high
temperature alarm is not triggered.

VISUAL CHECK OF GENERAL CONDITION AND STICKERS

The device should be checked for mechanical damage (general condition) and
for the completeness of sticker information, particularly the plate on the reverse.
There is a malfunction if mechanical damage to the device is evident which
could be harmful or impair the functional operation of the device.

ELECTRICAL SAFETY

All relevant electrical safety data should be checked, particularly the earth
conductor resistance (< 0.3 Ohm) and leakage current (< 0.75 mA). There is a
malfunction if there is a value outside the tolerances.

EN - Edition 03 / 2007

9

Summary of Contents for BW 585

Page 1: ...Instructions for use BW 585 Blood and Infusion Warmer 0123 EN Edition 03 2007...

Page 2: ...ork has been carried out BIEGLER Instructions for use BW 585 14 TECHNICAL DATA Device Blood and infusion warmer Type designation BW 585 Voltage 230 V 50 60 Hz or 110 V 50 Hz check the voltage specifie...

Page 3: ...C Values higher or lower than the ranges specified above may cause damage to the device or its accessories 16 EN Edition 03 2007 BIEGLER Instructions for use BW 585 Table of contents 1 Warnings and s...

Page 4: ...function deviating from the instructions for use do not use the device and return it to the service centre The periodic technical safety inspections must be carried out according to the section perio...

Page 5: ...ns Electromagnetic is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Page 6: ...w The customer or the user of the BW 585 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic dis...

Page 7: ...ns for use before putting the system into operation 3 1 SETTING UP PROCEDURE Fix the BIEGLER BW 585 firmly to the infusion stand using the clamps at the back Diag 1 9 Only use infusion stands or poles...

Page 8: ...is switched off To reset the device or switch off the alarm the device must first be disconnected from the power supply Attention The temperature alarm can also be set off externally e g by exposure t...

Page 9: ...The manufacturer s warranty will be void if it is found that the maintenance disinfection and inspection instructions have not been followed according to the instructions for use the device was damag...

Page 10: ...all the alarms are active but the heating is switched off The BW 585 now slowly cools down When the temperature drops below the 36 5 C threshold the low temperature alarm should be active For reasons...

Reviews: