Instructions for use - BW 585
BIEGLER
11 ELECTROMAGNETIC EMISSION
Table 201 – Guidance and manufacturer´s declaration – electromagnetic emission –
for all EQUIPMENT AND SYSTEMS (see 6.8.3.201 a) 3))
The BW 585 is intended for use in the electromagnetic environment specified below. The customer or the user
of the BW 585 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
The BW 585 uses RF energy only for its internal
function. Therefor, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
The BW 585 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
12
EN - Edition 03 / 2007
BIEGLER
Instructions for use - BW 585
3 INITIAL OPERATION
Users must familiarise themselves in detail with the contents of these
instructions for use before putting the system into operation.
3.1 SETTING UP PROCEDURE
Fix the BIEGLER BW 585 firmly to the infusion stand using the clamps at the
back (Diag.1/9). Only use infusion stands or poles that are sufficiently stable.
Connect power cable to power supply. Before connecting to mains power
supply, check the voltage specified on the device label. The device gives a short
beep and the standby light (Diag.1/5) lights up.
If a different temperature to 38.5°C is desired, it can be preset in Standby mode
using the controls
and
(Diag.1/6 and Diag.1/7). If an adjustment control is
pressed, the visual display indicates the existing preset temperature. By
repeated operation of the control
or
the temperature can be reset. The
indicator automatically goes out after approximately 7 seconds. Resetting of
temperature can only be performed in Standby mode.
Heating of the BW 585 can be started by pressing the control
(Diag.1/3).
The BW 585 attains the set target temperature within 1 minute. The indicator
display shows the actual temperature (+/- 0.5°C) as an illuminated band.
EN - Edition 03 / 2007
5
Diag. 1
1 Heat exchanger
2 Temperature scale
3 ON / STANDBY switch
4 LED indicator ON
5 LED indicator STANDBY
6 Control to decrease
temperature
7 Control to increase
temperature
8 Extension tube
9 Clamps
10 Entry of liquid
11 Exit of liquid