Instructions for use - BW 585

BIEGLER

11 ELECTROMAGNETIC EMISSION

Table 201 – Guidance and manufacturer´s declaration – electromagnetic emission – 

for all EQUIPMENT AND SYSTEMS (see 6.8.3.201 a) 3))

The BW 585 is intended for use in the electromagnetic environment specified below. The customer or the user
of the BW 585 should assure that it is used in such an environment.

Emissions test 

Compliance 

Electromagnetic environment - guidance 

RF emissions

CISPR 11 

Group 1 

RF emissions

CISPR 11

Class B 

Harmonic emissions

IEC 61000-3-2 

Class A 

Voltage fluctuations /
flicker emissions

IEC 61000-3-3

Complies 

The BW 585 uses RF energy only for its internal
function. Therefor, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.

The BW 585 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.

12

EN - Edition 03 / 2007

BIEGLER

Instructions for use - BW 585

3 INITIAL OPERATION

Users must familiarise themselves in detail with the contents of these
instructions for use before putting the system into operation.

3.1 SETTING UP PROCEDURE

Fix the BIEGLER BW 585 firmly to the infusion stand using the clamps at the
back (Diag.1/9). Only use infusion stands or poles that are sufficiently stable.

Connect power cable to power supply. Before connecting to mains power
supply, check the voltage specified on the device label. The device gives a short
beep and the standby light (Diag.1/5) lights up.

If a different temperature to 38.5°C is desired, it can be preset in Standby mode
using the controls 

 and 

 (Diag.1/6 and Diag.1/7). If an adjustment control is

pressed, the visual display indicates the existing preset temperature. By
repeated operation of the control 

 or 

 the temperature can be reset. The

indicator automatically goes out after approximately 7 seconds. Resetting of
temperature can only be performed in Standby mode.

Heating of the BW 585 can be started by pressing the control 

 (Diag.1/3). 

The BW 585 attains the set target temperature within 1 minute. The indicator
display shows the actual temperature (+/- 0.5°C) as an illuminated band.

EN - Edition 03 / 2007

5

Diag. 1

1 Heat exchanger
2 Temperature scale
3 ON / STANDBY switch
4 LED  indicator  ON
5 LED indicator STANDBY
6 Control to decrease

temperature

7 Control to increase

temperature

8 Extension tube
9 Clamps
10 Entry of liquid
11 Exit of liquid

 Flow direction

Summary of Contents for BW 585

Page 1: ...Instructions for use BW 585 Blood and Infusion Warmer 0123 EN Edition 03 2007...

Page 2: ...ork has been carried out BIEGLER Instructions for use BW 585 14 TECHNICAL DATA Device Blood and infusion warmer Type designation BW 585 Voltage 230 V 50 60 Hz or 110 V 50 Hz check the voltage specifie...

Page 3: ...C Values higher or lower than the ranges specified above may cause damage to the device or its accessories 16 EN Edition 03 2007 BIEGLER Instructions for use BW 585 Table of contents 1 Warnings and s...

Page 4: ...function deviating from the instructions for use do not use the device and return it to the service centre The periodic technical safety inspections must be carried out according to the section perio...

Page 5: ...ns Electromagnetic is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Page 6: ...w The customer or the user of the BW 585 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic dis...

Page 7: ...ns for use before putting the system into operation 3 1 SETTING UP PROCEDURE Fix the BIEGLER BW 585 firmly to the infusion stand using the clamps at the back Diag 1 9 Only use infusion stands or poles...

Page 8: ...is switched off To reset the device or switch off the alarm the device must first be disconnected from the power supply Attention The temperature alarm can also be set off externally e g by exposure t...

Page 9: ...The manufacturer s warranty will be void if it is found that the maintenance disinfection and inspection instructions have not been followed according to the instructions for use the device was damag...

Page 10: ...all the alarms are active but the heating is switched off The BW 585 now slowly cools down When the temperature drops below the 36 5 C threshold the low temperature alarm should be active For reasons...