Instructions for use - BW 585
BIEGLER
1 WARNINGS AND SAFETY INSTRUCTIONS
•
In the event of any suspected malfunction while in operation, the device
should be immediately removed from service and not used for infusions or
transfusions until appropriate investigations have demonstrated that there has
been no impairment.
•
If the high temperatur alarm is triggered the supply of liquid to the patient must
be immediately stopped by disconnecting the connection tube to the patient.
The medium being used in the device must no longer be administered to the
patient.
•
Non-compliance with the flow direction specified in the instruction manual may
lead to a deviation in the selected control temperature. This could result in
damage to the medium.
•
The device may only be fastened to infusion stands or tripods which are
suitable due to their stability and load capacity.
•
Only sterile BIEGLER consumable materials may be used in conjunction with
the BW 585.
•
The device must only be used in areas in which the electrical installations are
in accordance with the rules and regulations in force.
•
Safe disconnection from the mains power supply can only be achieved by
unplugging the mains plug.
•
The device must not be used in rooms subject to explosion hazard.
•
Repairs and modifications to the device may only be carried out by persons or
service centres authorized by BIEGLER.
•
The device must not be immersed in liquids or sterilized with steam or by
thermochemical methods.
•
All extraneous influences such as radiation or high temperatures are to be
kept to a minimum.
•
Avoid exerting force to the device or its accessories.
•
If the device is dropped, damaged by force or if it shows a function deviating
from the instructions for use, do not use the device and return it to the service
centre.
•
The periodic technical safety inspections must be carried out according to the
section „periodic inspections“.
2
EN - Edition 03 / 2007
BIEGLER
Instructions for use - BW 585
Table 206 – Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM -
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING (see 6.8.3.201 b))
Recommended separation distances between
portable and mobile RF communications equipment and the BW 585
The BW 585 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the BW 585 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the BW 585 as recommended below, according to the maximum output power of the communications
equipment.
11.1.1.1 Separation distance according to frequency of transmitter
m
Rated maximum output of
transmitter
W
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0,12
0,04
0,07
0,1
0,37
0,11
0,22
1
1,17
0,35
0,70
10
3,69
1,11
2,21
100
11,67
3,50
7,00
For transmitters rated at a maximum output power not listed above the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
EN - Edition 03 / 2007
15
P
d
17
,
1
=
P
d
35
,
0
=
P
d
7
,
0
=