Instructions for use - BW 585
BIEGLER
Table 204 – Guidance and manufacturer´s declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING (see 6.8.3.201 b))
Guidance and manufacturer´s declaration – electromagnetic immunity
The BW 585 is intended for use in the electromagnetic environment specified below. The customer
or the user of the BW 585 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
150 kHz to 80 Mhz
3 V/m
80 MHz to 2,5 GHz
Portable and mobile RF communications equipment
should be used no closer to any part of the BW 585,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
11.1.1 Recommended separation distance
80 MHz to 800 Mhz
800 MHz to 2,5 GHz
where
p
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in metres (m).
b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the BW 585 is used exceeds the applicable RF compliance level above, the BW 585 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the BW 585.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
14
EN - Edition 03 / 2007
P
d
17
,
1
=
P
d
35
,
0
=
P
d
7
,
0
=
BIEGLER
Instructions for use - BW 585
The BW 585 must not be used in the following circumstances:
•
If the case is damaged or one of the front film layers becomes detached
•
If the device has been exposed to a hard physical shock (e.g. dropped, hit or
shaken)
•
If the device has been immersed in water
•
If the device has triggered a high temperatur alarm that was not caused by
external factors
•
If the mains power cord or plug is damaged
•
If the device has given somebody an electric shock
•
If the fixing clamps are damaged and no longer assure safe clamping to the
infusion stand
Should a malfunction be evident, suitable warning signs should be attached to
the device to ensure that it cannot be used before necessary service and repair
work has been carried out.
EN - Edition 03 / 2007
3