Instructions for use - BW 585

BIEGLER

Table 204 – Guidance and manufacturer´s declaration – electromagnetic immunity – 

for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING  (see 6.8.3.201 b))

Guidance and manufacturer´s declaration – electromagnetic immunity 

The BW 585 is intended for use in the electromagnetic environment specified below. The customer
or the user of the BW 585 should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level

Compliance

level

Electromagnetic environment -
guidance 

 

 

 

 

Conducted RF

IEC 61000-4-6

 

 

Radiated RF

 

IEC 61000-4-3

 

3 V

150 kHz to 80 Mhz

3 V/m

 

80 MHz to 2,5 GHz

 

Portable and mobile RF communications equipment
should be used no closer to any part of the BW 585,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.

11.1.1 Recommended separation distance

       

80 MHz to 800 Mhz

     

  800 MHz to 2,5 GHz

where 

p

 is the maximum output power rating of the

transmitter in watts (W) according to the transmitter
manufacturer and 

d

 is the recommended separation

distance in metres (m).

b

 

Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,

a

should be less than the compliance level in each
frequency range.

b

 

Interference may occur in the vicinity of equipment
marked with the following symbol:

NOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2  These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, objects
and people.

a

      Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur

radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the BW 585 is used exceeds the applicable RF compliance level above, the BW 585 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the BW 585.

b

      Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

14

EN - Edition 03 / 2007

P

d

17

,

1

=

P

d

35

,

0

=

P

d

7

,

0

=

BIEGLER

Instructions for use - BW 585

The BW 585 must not be used in the following circumstances:

If the case is damaged or one of the front film layers becomes detached

If the device has been exposed to a hard physical shock (e.g. dropped, hit or
shaken)

If the device has been immersed in water

If the device has triggered a high temperatur alarm that was not caused by
external factors

If the mains power cord or plug is damaged

If the device has given somebody an electric shock

If the fixing clamps are damaged and no longer assure safe clamping to the
infusion stand

Should a malfunction be evident, suitable warning signs should be attached to
the device to ensure that it cannot be used before necessary service and repair
work has been carried out.

EN - Edition 03 / 2007

3

Summary of Contents for BW 585

Page 1: ...Instructions for use BW 585 Blood and Infusion Warmer 0123 EN Edition 03 2007...

Page 2: ...ork has been carried out BIEGLER Instructions for use BW 585 14 TECHNICAL DATA Device Blood and infusion warmer Type designation BW 585 Voltage 230 V 50 60 Hz or 110 V 50 Hz check the voltage specifie...

Page 3: ...C Values higher or lower than the ranges specified above may cause damage to the device or its accessories 16 EN Edition 03 2007 BIEGLER Instructions for use BW 585 Table of contents 1 Warnings and s...

Page 4: ...function deviating from the instructions for use do not use the device and return it to the service centre The periodic technical safety inspections must be carried out according to the section perio...

Page 5: ...ns Electromagnetic is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Page 6: ...w The customer or the user of the BW 585 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic dis...

Page 7: ...ns for use before putting the system into operation 3 1 SETTING UP PROCEDURE Fix the BIEGLER BW 585 firmly to the infusion stand using the clamps at the back Diag 1 9 Only use infusion stands or poles...

Page 8: ...is switched off To reset the device or switch off the alarm the device must first be disconnected from the power supply Attention The temperature alarm can also be set off externally e g by exposure t...

Page 9: ...The manufacturer s warranty will be void if it is found that the maintenance disinfection and inspection instructions have not been followed according to the instructions for use the device was damag...

Page 10: ...all the alarms are active but the heating is switched off The BW 585 now slowly cools down When the temperature drops below the 36 5 C threshold the low temperature alarm should be active For reasons...

Reviews: