Instructions for use - BW 585
BIEGLER
Select suitable consumable material. See section „consumable material“.
Prepare infusion or transfusion. Attention: The length of the tube between the
BW 585 and the patient must be at least 40 cm and the tube must not be
streched.
Beginning at the back of the heat exchanger the extension tube is gently pulled
and coiled in an anti-clockwise direction towards the front. It is advisable that the
gap between the BW 585 and patient does not exceed much more than 80 cm.
Important: The tube must be completely inserted into the groove (Diag. 2).
The flow direction specified in diagram 1 must be complied with.
3.2 ALARMS
The BW 585 can trigger two types of temperature alarm:
The low temperature alarm is activated when the temperature of the heat
exchanger drops below 36.5°C. The audio low temperature alarm is deactivated
during the first 60 seconds after switching on.
The high temperature alarm is activated when the temperature of the heat
exchanger exceeds 42.0°C. In this event visual and audio alarm signals are
given and the heating is switched off. To reset the device or switch off the alarm,
the device must first be disconnected from the power supply.
Attention: The temperature alarm can also be set off externally e.g. by exposure
to sunlight.
3.3 SHUTTING DOWN THE DEVICE
After being used for treatment, the device is shut down as follows:
6
EN - Edition 03 / 2007
Diag. 2
Position of the tube in the
groove of the heat exchanger
BIEGLER
Instructions for use - BW 585
If the manufacturer is required to meet a warranty claim in accordance with
these terms, the customer shall bear the costs and risks of transport of the
device from and to the place of use.
The manufacturer and/or supplier under no circumstances assume liability for
slight negligence. The compensation for lost earnings and profits is likewise
excluded.
9 RETURN OF DEVICES
Devices must be carefully cleaned and disinfected before being placed in the
original packaging for returning.
If the original packaging is no longer available, the product has to be suitably
packaged for the method of dispatch.
10 MANUFACTURER`S DECLARATION
The blood and infusion warmer BW 585 is a medical product as defined by
Directive 93/42/EEC.
This is documented with the CE mark.
Notified Body: TÜV Product-Service, Approval Number CE0123
EN - Edition 03 / 2007
11