14 TECHNICAL DATA
Device:
Blood and infusion warmer
Type designation:
BW 585
Voltage:
230 V / 50-60 Hz or 110 V / 50 Hz
check the voltage specified on the device label
Power consumption:
280 W
Type of protection
against electric shock:
I
Degree of protection
against electric shock:
B
Degree of protection
against ingress of liquids:
IPX4
Safety protection:
230V: primary 2 x 1.6 AT secondary 315 mAT
110V: primary 2 x 3.15 AT secondary 315 mAT
Temperature setting:
37°C - 41°C adjustable in steps of 0.5°C
High temperature cut-out:
42°C / 42.5°C / 45°C±3°C
Max. system pressure:
300 mmHg
Dimensions:
WxHxD 140 x 190 x 240 mm
Weight:
2.5 kg
Classification:
IIb according to rule 9
Type of operation:
continuous operation