26
Problem
Cause
Solution
The power is unable to
turn on.
Battery has run out.
Recharge the battery.
No battery inserted.
Insert battery.
The internal fuse has burnt.
Contact your dealer.
Charge failure
Battery has been over used
Replace with new battery.
The power cord plug is not
inserted into the charger
base well
plug the power cord or
adaptor properly
The handpiece is not
correctly match to the
charger base
Reload the handpiece to get
a good match
There is some residual
Contamination on the
charger base
remove the residual
contamination
An error code is displayed.
Refer to No. 10 Error Code.
Nothing is displayed on the
panel of the handpiece even
when it is loaded to the
charger
Contact your dealer.
The charger base is broken
Charge the motor by cable,
then contact your dealer
The motor hand piece
does not rotate.
The contra angle has been
blocked
Clean or reset the contra
angle
The alarm sounds when
the motor is working
The contra angle head has
jammed up.
Contra rotation failure
Set to Auto Reverse mode,
take off contra angle, there
is NO click sound when
working
Contact your dealer.
Set to Auto Reverse mode,
take off contra angle, there
is click sound
Clean the contra angle
The alarm sounds when
the motor is working.
There is some residual
contamination on the
rotating shaft of the contra
angle
Clean the contra angle