25
10. Error Code
11. Troubleshooting
Error ode
Error
Cause
Check Remedy
When
working
E-0
Self-Check
error
Malfunction of circuit
Contact your dealer
E-1
Overcurrent
The motor handpiece
is locked. (auto
reverse mode)
Remove payload
E-2
Overvoltage
Malfunction of circuit
Contact your dealer
When
Charging
E-3
Charger
failure
Malfunction of circuit
Contact your dealer
E-4
Low voltage
of battery
Battery out of service
Chang Battery
E-5
High voltage
of battery
Battery out of service
Chang Battery
E-8
Low voltage
of adaptor
Battery out of service
Change adaptor
E-9
High voltage
of adaptor
Battery out of service
Change adaptor
When
calibration
E-7
Calibration
failure
Motor damaged or
handpiece blocked
Replace the motor
handpiece or motor
If something goes wrong, please check the following again before consulting your
dealer. If none of these is applicable or the trouble is not remedied even after action
has been taken, a failure of this product is suspected. Contact your dealer, do not
field-repairable.