22
7. Cleaning Maintenance
• Insert the spray nozzle into the contra
angle head(insert to the part connect
to the main unit)
• Insert the lubricants spray nozzle to
the provided spray nozzle loop, inject
the lubricants for 2-3 seconds till
the outlet liquid from the head of the
contra angle clean
7.1 Lubricatinq Contra AngIe
• Use qualified SPRAY to lubricate or sterilization. " Lubricate the contra angle only.
“Apply SPRAY/SPRAY Plus after each use and/or before each calibration and
autoclaving
7.2 Cleaning the Motor Handpiece
• Wipe it off with a cotton cloth moistened with alcohol.
7.3 Cleaning charger
• Wipe it off with a cotton cloth moistened with alcohol.
Warning
• Do not lubricate the motor handpiece or the charger.
• Before mounting the lubricated contra angle to the motor handpiece, wipe off
excess oil. Stand it on its end or lean it in the proper position for gravity draining.
Mount it after excess oil has been completely drained.
• Hold the contra angle securely to prevent it from flying off by the pressure of the
spray.
• Hold the SPRAY (option) upright.
• When cleaning the motor headpiece, do not use any solvent such as benzene
and thinner.