Problem: The test does not start after inserting the test strip into the monitor
and applying the blood.
Cause
Remedy
Blood volume too low
Repeat test with new test strip and greater
volume of blood.
Faulty test strip
Repeat test with a new test strip.
Blood was applied with the
monitor switched off
Repeat test, apply blood when
6
flashes.
Faulty coding
Repeat coding (see "6.1 Checking code number
(coding)“ page 24).
Basic settings of monitor
have been changed and the
change was not completed
(see "4.2 Making basic
settings“ page 18).
Press 'SET' button until OFF is displayed.
Repeat test.
Faulty monitor
Contact Customer Service.
Problem: The blood pressure value is unusually high or low
Cause
Remedy
You moved or talked during
the measurement.
You should wait five minutes before measuring
a second time.
Repeat the test.
Repeat test and do not move or talk during the
test. Make sure your body position is correct
during the test.
You are not yet relaxed
enough.
Rest for approx. 10 minutes.
Repeat the test.