Beurer BGL40 mg/dL
42
No.
Cause
Remedy
Err 08
Cuff pressure out of range
Repeat test and do not move or talk
during the test.
Err 09
Cuff pressure is insufficient
even at maximum pump
pressure
The blood pressure exceeds the
measurement range. Repeat
measurement. if the problem persists,
seek medical advice.
Err 17
No pulse detectable
Repeat measurement. if the problem
persists, seek medical advice.
Err 32
Error! Too much
movement during
measurement
Do not move or speak during
measurement.
Problem: The monitor does not switch on.
Cause
Remedy
Low batteries
Replace batteries.
Battery missing or not
installed correctly
Check that the batteries are correctly installed (see
"4.1 Inserting and replacing batteries“ page 18).
Test strip is
incompletely inserted
or wrong way around
Insert a test strip into the monitor with the contacts
first. Make sure that the front of the test strip is facing
you (see "3.3 Test strips" page 16).
Faulty monitor
Contact Customer Service.