Better Water LLC; rev. May 2020
Page 54 of 78
1232 RO Operator Manual
PREPARATION
1.
Make sure that there are no requirements for RO water (
no patients are dialyzing
).
2.
Turn the RO OFF and unplug or disconnect from the main power source.
3
. Turn OFF the
RO Inlet Valve from the Pre-treatment
.
4.
Depress the red pressure relief button on the Big Blue Pre-Filter to the RO to relieve the
pressure.
UNPACK MEMBRANES
5.
Open and unpack the membranes from their shipping cartons.
6. Verify Correct Membranes
- Before opening the boxes of the new membranes, verify that they are the correct membranes
for your RO.
- If they are not the correct membranes, DO NOT ATTEMPT to install them.
- A
“
RO MEMBRANE
” label should be adhered to the box. This label will have a received date.
Membrane is good for one year from this date.
- Each membrane is also labeled with an individual serial number which should be recorded and
also noted in which membrane housing the specific membrane will be installed.
7. Open and Unpack the New Membrane
- Open the membrane box.
- Remove the bagged membrane from the box discarding the packing material.
- Do not open the bag containing the membrane
REPLACE MEMBRANES
8.
Remove the 3/8" Blue Product Tubing from the 3/8" x 1/4" Fast-N-Tite.
9.
Remove the 3/8" x 1/4" Fast-N-Tite.
10.
Remove the U-Pin
11.
Thread the tee-handle tool into the Upper End-Cap until snug (
do not over tighten
)
and pull up to remove it.
12.
Using channel locks or a similar tool to grasp the Membrane, pulling up to remove it.
13.
Remove the plug that is screwed into the Lower End-Cap.
14.
Thread the tee-handle tool into the Lower End-Cap until snug (do not over tighten)
and pull down to remove it.
15.
Lubricate the O-Rings in both End-Caps with medical grade glycerin.
- Glycerin is recommended, KY Jelly can also be used
-
DO NOT USE VASELINE, PLUMBER’S GREASE OR SILICONE GREASE.
16.
Replace the Lower End-Cap, tightening with the tee-handle tool (
do not overtighten
).
17.
Replace the plug in the Lower End-Cap.
18.
Be sure to wear sterile gloves when handling the new membranes.
- CAUTION: As stated before, the new membrane is packed in a liquid preservative, so take
caution in handling to prevent slippage and exposure to prevent injury.
Summary of Contents for 1232 RO
Page 1: ...REA 2390 1232 RO Operator Manual Better Water LLC rev May 2020 ...
Page 8: ...Better Water LLC rev May 2020 Page 4 of 78 1232 RO Operator Manual ...
Page 15: ...Better Water LLC rev May 2020 Page 11 of 78 1232 RO Operator Manual DEVICE INFORMATION ...
Page 47: ...Better Water LLC rev May 2020 Page 43 of 78 1232 RO Operator Manual SYSTEM MAINTENANCE ...
Page 74: ...Better Water LLC rev May 2020 Page 70 of 78 1232 RO Operator Manual ...
Page 80: ...Better Water LLC rev May 2020 Page 76 of 78 1232 RO Operator Manual ...
Page 82: ...Better Water LLC rev May 2020 Page 78 of 78 1232 RO Operator Manual ...