BRINGING THE RO BACK FROM STORAGE
*
1.
Remove protective plastic bags from openings.
2.
Reconnect to drain.
3.
Reconnect to water source.
4.
Run product hose to drain.
5.
Reconnect to power.
6.
Install a new Pre-Filter, and reinstall the Pre-Filter Housing.
7.
Remove the Final-Filter Housing, taking caution that it will be full of preservative which should
be properly discarded. Remove the false-tube, install a new Final-Filter, and then reinstall the
Final-Filter Housing.
8.
Rinse for at least two hours.
9.
Clean and disinfect the RO.
10.
Sample as per your procedures for colony count (cultures), endotoxins (LAL), and AAMI.
WARNING
Preservatives can cause serious injury or death.
Proper protective equipment must be used.
The preparation and handling of these chemical solutions must be done in accordance
with the specifications established for the particular chemical and their Material Safety
Data Sheet (MSDS).
These procedures should be performed by trained and qualified technicians.
Use extreme care when performing this procedure to prevent exposure to the
preservative solution inside the RO. Use caution when removing bags from covered
ports, and opening valves or housings.
Summary of Contents for 1232 RO
Page 1: ...REA 2390 1232 RO Operator Manual Better Water LLC rev May 2020 ...
Page 8: ...Better Water LLC rev May 2020 Page 4 of 78 1232 RO Operator Manual ...
Page 15: ...Better Water LLC rev May 2020 Page 11 of 78 1232 RO Operator Manual DEVICE INFORMATION ...
Page 47: ...Better Water LLC rev May 2020 Page 43 of 78 1232 RO Operator Manual SYSTEM MAINTENANCE ...
Page 74: ...Better Water LLC rev May 2020 Page 70 of 78 1232 RO Operator Manual ...
Page 80: ...Better Water LLC rev May 2020 Page 76 of 78 1232 RO Operator Manual ...
Page 82: ...Better Water LLC rev May 2020 Page 78 of 78 1232 RO Operator Manual ...