5 Electric Installation
LB 470 Level
46
56925BA2 Rev. 03, 11/2017
4.
Open the terminal connection (Fig. 21, item 3) with an operating tool (slot-
ted screwdriver) and insert the stripped wire (min. 8 mm) (Fig. 21,item 4). The
terminal connection closes by pulling out the operating tool.
5.
The terminal connections are designed for wires (flexible) with a conductor
cross-section from 0.2 mm² to 2.5 mm² (with ferrule without plastic sleeve) or
0.2 mm² to 1.5 mm² (with ferrule with plastic sleeve). AWG = 24 – 12.
6.
Plug the network plug into the RJ45 socket (Fig. 21, item 9) (optional).
7.
Check the correct connection of the PE conductor (Fig. 21, item 10).
NOTICE
Note the specification relating to Protective earth and equipotential bonding
in chapter 5.1.4.
NOTICE
Only cable that is suitable for connection to the corresponding terminals may
be used. Detailed specifications can be found in the chapter 5.1.2 Cables and
Lines.