LB 470 Level
7 Main Menu Device Setup
56925BA2 Rev. 03, 11/2017
137
Analogue output: AO Monitoring
Device settings | Setup | Output | Analogue Output (AO) | AO Monitoring
If
″
AO Monitoring" is activated (Fig. 120, item 1), the current output will be moni-
tored. It is continuously monitored whether the current value flowing in the
current loop is correct.
In the event of a variation e.g. owing to an error in the hardware, too large a load
or a disruption in the loop, an error message is triggered.
1
2
3
Activate check box "AO Monitoring Enabled"
Target value [mA]
Actual value [mA]
Fig. 120 Analog output monitoring