LB 470 Level
4 Installation
56925BA2 Rev. 03, 11/2017
37
Installed with terminal panels
j
DANGER
Danger to life from electric shock!
Installation/maintenance may only be carried out if the device has been
de-energised.
Test of absence of harmful voltages when the front side is open.
In case of an electric shock, carry out first aid measures and immediately call an
emergency service.
NOTICE
The master's EVUs / slave modules must be secured against pulling out by fix-
ing screws (Fig. 19, item 5).
The device must be disconnected from the mains voltage before it is pulled
out.
1
2
3
Master EVU
Socket board terminal panel
Guide rails
4
5
Drilling
Fixing screws
Fig. 19
19" subrack with terminal panel (Ex.: 4x master)
1.
Set master EVU / slave module (Fig. 19, item 1) into the guide rails (Fig. 19,
item 3).
2.
Carefully slide module into the subrack until the plug connector is inserted
into the socket board (Fig. 19, item 2).
3.
Tighten fixing screws (Fig. 19, item 5).
Summary of Contents for Duo Xpert
Page 2: ......
Page 4: ......
Page 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Page 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Page 184: ......
Page 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Page 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Page 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...