LB 470 Level
7 Main Menu Device Setup
56925BA2 Rev. 03, 11/2017
123
Calculate Factor M
The "Factor M" (Fig. 102, item 2) is close to 1 (default value) for standard applica-
tions.
1.
Click the button
< Calculate >
(Fig. 102, item 3) to determine Factor M from
the count rate.
Factor M has been calculated and is displayed in the input field (Fig. 102,
item 2).
2.
Click the button
< Close
>
(Fig. 102, item 6) to return to
″
GPC calibration
″
in
the submenu.
The determined Factor M has been accepted.
IMPORTANT
The M factor can be entered manually (Fig. 102, item 2), if a determination of
the counting rates is not possible.
GPC calibration: Max. GPC Factor
The maximum factor with which the level count rate may be compensated is spec-
ified in the field "Max. GPC Factor" (Fig. 99, item 5). The default value is 3.0.
GPC settings
Device settings | Setup | Calibration | GPC | GPC settings
The isotope specified by the source can be selected for the gas density in the sub-
menu
″
GPC settings
″
(Fig. 103, item 1). The half-life of the isotope is shown on the
display field (Fig. 103, item 2).
NOTICE
The sources for level measurement and the sources of the GPC measurement
must contain the same Nuclide.
1
2
Selection arrow "Nuclide"
Display field
″
Half-life
″
[years]
Fig. 103 Submenu "GPC settings"
Summary of Contents for Duo Xpert
Page 2: ......
Page 4: ......
Page 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Page 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Page 184: ......
Page 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Page 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Page 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...