![Berthold Duo Xpert Operating Manual Download Page 142](http://html1.mh-extra.com/html/berthold/duo-xpert/duo-xpert_operating-manual_2739249142.webp)
7 Main Menu Device Setup
LB 470 Level
142
56925BA2 Rev. 03, 11/2017
Digital Outputs (DO)
Device settings | Setup | Output | Digital Outputs (DO)
The signals of the digital outputs are switched via potential-free relay contacts. The
contacts are controlled "fail safe", ie, in the event of an alarm, the current at the
relay coil drops and the NO contact (normally open) is opened. The wiring diagrams
in chapter 5 show the relay contacts in the de-energized state.
1
2
3
State DO-1 (ACTIVE / PASSIVE)
State DO-2 (ACTIVE / PASSIVE)
Selection arrow Function DO-2
4
5
6
State DO-3 (ACTIVE / PASSIVE)
Selection arrow Function DO-3
Function selection
Fig. 124 Digital Outputs
The alarm relays 1 and 2 can be assigned to the following functions in the event of
an alarm:
PV Low alarm
The relay switches when the value at
Device
Setup | Setup | Alarms | PV alarm settings
is
underrun.
PV High alarm
The relay switches if the value under
Device
Setup | Setup | Alarms | PV Alarm Settings
is
exceeded.
Detector temperature Alarm
The relay switches when values set at
Device
Setup | Setup | Alarms | Det.-Temp. Alarm
function
are exceeded or underrun.
XIP Alarm
The relay switches when detection is acti-
vated at
Device settings | Setup | Signal
processing | XIP
and interference was de-
tected.
Source Replacement Alarm
The relay switches when notification at
De-
vice settings | Setup | Signal processing |
Source replacement
is activated and interfer-
ence is detected.
Measurement Stopped
The relay switches on during tests or other
states where the measurement is stopped.
For example, Simulation, plate space meas-
urement, and detector update.
System State: Warning
The relay switches when the event message
″
Warning
″
is displayed.
Summary of Contents for Duo Xpert
Page 2: ......
Page 4: ......
Page 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Page 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Page 184: ......
Page 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Page 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Page 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...