LB 470 Level
7 Main Menu Device Setup
56925BA2 Rev. 03, 11/2017
103
IMPORTANT
All entries and changes in the
″
Parameters
″
tab will take effect only when you
click on the
< Calibrate >
button click in the Calibrate Submenu.
1
2
3
4
5
6
Input field Background [cps]
Button
<Ok>
Input field count rate
Input field measuring Duration
Button
<Start>
Button
<Cancel>
Fig. 82
Calibration parameters: Background
Determine background
Device settings | Setup | Calibration | Calibration Settings
1.
Click on the text field "Background" (Fig. 82, item 1).
A new window "Background" opens to determine the background Count
rate.
NOTICE
Influences from neighbouring sources must be excluded in order to avoid errors
in the measurement of natural background radiation.
2.
Click on the
″
Read-In-Time
″
field and specify the duration of the measure-
ment in seconds. The higher you set the measurement time, the more
accurate the result.
3.
Confirm with the Enter key and click on the button
<Start>
to start the meas-
urement.
The measurement is performed.
4.
Click on
<OK>
to accept the count rate.
Summary of Contents for Duo Xpert
Page 2: ......
Page 4: ......
Page 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Page 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Page 184: ......
Page 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Page 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Page 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...