8.8
Problems, causes and solutions
The
BENOMIC S500
is equipped with various safety systems that can temporarily block
the intended functions, for example through incorrect operation.
In order to alert the user to a changing status or incorrect operation of the
BENOMIC
S500
during use, a multi-colour indicator (11) and horn is used. For the situations below,
first always refer to chapters 5.3.1 and 5.3.2!
Problem A :
The BENOMIC S500 does not run.
Cause
:
Key switch is disabled
Solution
:
Enable key switch (vertical)
Emergency stop locked
Release emergency stop (turn / pull out)
Speed potentiometer at 0
Set a speed
Motor controller fault
Switch off and on again with the main switch
Consult your dealer
Battery terminals make poor contact
Clean the battery terminals, remount the clamps
Defective foot pedal
Replace foot pedal, consult your dealer
Lifter not withdrawn far enough or sensor defective
Withdraw lifter (system) fully (12) or check sensor
Stabilizers are not fully retracted or sensor defective
Retract stabilizers fully (left & right) (8.4) or check sensor
25A circuit breaker in 0-position
Consult your dealer
6.3 A control circuit fuse defective.
Consult your dealer
Problem B :
Speed can be poorly adjusted.
Cause B :
Speed control button is defective.
Solution
:
Consult your dealer
Pipe-rail sensor is defective (only runs slowly)
Consult your dealer
Drive motor control is defective
Consult your dealer