During commissioning it is recommend
that initially the second digit of the rate
parameters is set to 0 (step response
off) and the first digit (amount of
filtering) is adjusted to provide
acceptable rate display stability. The
second digit should them be increased
until acceptable rate display stability is
once again achieved.

To adjust the filter parameters select
‘FILtEr’ from the main menu and press

P

 to reveal the current setting.  The

first digit will be flashing and may be
adjusted using the 

Up

 or 

Down 

button.

The 

P

 button will transfer control to the

second digit.  When both are set as
required, return to the ‘FILtEr’ prompt
in the main menu by pressing 

E

.

   Note:

  While adjusting the filter, the rate is

shown on the lower display so that
stability can be assessed.

Step 8    Enter the total scaling factor

The K factor of the flowmeter in this
example is 1050.0 pulses per litre and
the total display is required in cubic
metres.

There are 1000 litres in a cubic metre
so the BA534D is required to display 1
cubic metre for every 1,050 x 1,000
pulses received.  ‘SCALE-t’ should
therefore be set to 1,050,000

Select ‘SCALE-t’ from the main menu
and press 

P

 which will reveal the

current setting with one digit flashing.
To position the decimal point, operate
the 

P

 push-button to move the flashing

digit to the least significant position.
Pressing the 

Up

 and 

Down

 buttons

simultaneously will then remove the
decimal point.

Using the 

Up

 and 

Down

 buttons to

adjust each digit in turn and the 

P

button to transfer control between
digits, enter 1050000    Finally return to
the 'SCALE-t' prompt in the main menu
by pressing 

E

.

Step 9

Define clip-off

In this example totalisation is to be
inhibited at flow rates below 300 litres
per hour.

Select ‘CLIP-OFF’ from the main menu
and press 

P

 which will reveal the clip-

off threshold in litres per hour i.e. the
same units already selected for the
rate display.  The most significant digit
of the display will be flashing,
indicating that this digit may be
adjusted.    Pressing 

five times will

move the flashing digit to the third
lease significant digit.

Using the 

Up

 and 

Down

 buttons set

the third least significant digit to 3  i.e.
a clip-off threshold of 300.     Finally
return to the 'CLIP-OFF' prompt in the
main menu by pressing 

E

.

Step 10 Turn local reset on

In this example the operator needs to
reset the total display to zero by
pressing the 

Up

 and 

Down

 buttons

simultaneously.  To activate the local
reset function select ‘LOC.rSEt’ from
the main menu and press 

P

 to display

the current setting.  Pressing the 

Up

 or

Down

 button will toggle the display

between ‘On’ and ‘OFF’.  Select ‘On’
and press 

E

 to return to the ‘LOCrSEt’

prompt in the main menu.

Step 11 Reset the grand total to zero

Before completing commissioning  the
grand total counter should be reset to
zero.  Select 'CLr. Gtot' from the main
menu and press

 P

 which will cause the

instrument to display 'Clr. no'.   Toggle
this to 'CLr. YES' using the 

Up

 or

Down

 push-buttons and press 

P

 which

will result in a '0000' prompt.  Using the

Up

, and

 Down

 buttons and the

  P

button to move to the next digit,
confirm the request by entering the
password 'SurE'. Note: S is entered as
5.    Pressing

  E

 will then reset the

grand total counter to zero and return
the instrument to the 'CLr. Gtot' prompt
in the main menu.

Summary of Contents for BA534D

Page 1: ...Issue 10 18th September 2008 BA534D Externally powered pulse input field mounting rate totaliser issue 10...

Page 2: ...Calibration procedure 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comments 8 Accessorie...

Page 3: ...ys further extend the applications Fig 1 Simplified block diagram of BA534D 2 1 Initalisation Each time power is applied to the instrument initalisation is performed After a short delay the following...

Page 4: ...et Activated when the total Annunciator display is being reset 3 SYSTEM DESIGN 3 1 Flow transmitter loops The BA534D may be connected to almost any pulse output flowmeter and calibrated to display rat...

Page 5: ...4D rate totaliser is housed in a robust IP66 glass reinforced polyester GRP enclosure incorporating an armoured glass window and stainless steel fittings It is suitable for exterior mounting in most i...

Page 6: ...rective 2004 108 EEC For specified immunity all wiring should be in screened twisted pairs and the BA534D earth terminal should be locally earthed or connected to an earthed cable screen To prevent ci...

Page 7: ...total scale factor SCALE t Total flow is continuously added to the grand total The BA534D displays real decimal points Moving the position of a decimal point in a scale factor will therefore affect th...

Page 8: ...8...

Page 9: ...2 Lower display Turns the lower display which normally shows rate on or off See section 5 7 Display Description of function SCALE r Rate Scale Factor Defines the arithmetic relationship between the p...

Page 10: ...s from a wide variety of sensors To define an input type select InPut from the menu and press P which will reveal the current input type Pressing Up or Down will scroll through the five options UOLts...

Page 11: ...igit can be changed by pressing the Up or Down buttons When this digit is correct pressing P will transfer control to the next digit To position the decimal point in the rate scale factor move the fla...

Page 12: ...r control to the next digit To position the decimal point in the scaling factor move the flashing digit to the left hand side of the required decimal point position and simultaneously press the Up and...

Page 13: ...git of the code can be changed using the Up and Down push buttons and the P button to move to the next digit When the required code has been entered press E to return to the main menu The revised secu...

Page 14: ...ush button until the third least significant digit of the total display and following decimal point are flashing This gives the required total display resolution of 0 01 Press P to move control to the...

Page 15: ...enter 1050000 Finally return to the SCALE t prompt in the main menu by pressing E Step 9 Define clip off In this example totalisation is to be inhibited at flow rates below 300 litres per hour Select...

Page 16: ...ode 1209 digit by digit The P button will transfer control between digits When the new code has been entered press E to return to the main programme menu Step 13 Return to the display mode Following c...

Page 17: ...20mA output calibration Alarms do not function Alarms have been disabled following calibration change Re enable both alarms 7 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING M...

Page 18: ...utputs WARNING These alarm outputs should not be used for critical safety applications such as a shut down system When the BA534D power supply is turned off or disconnected both BA534D alarm outputs w...

Page 19: ...19 Fig 10 Alarm menu structure...

Page 20: ...available on rate alarms See section 8 2 8 Summary of programmable alarm functions dELA Alarm delay time Adjusts the delay between the display equalling the setpoint and the alarm output being activa...

Page 21: ...st the hysteresis select HStr from the alarm menu and press P which will reveal the existing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next d...

Page 22: ...o the setpoints in the display mode by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with the security code set to 0000 8 2 12 Adjusting alarm setpoints from th...

Page 23: ...pulse output The output is a polarised current sink that closes for a programmable duration each time the least significant digit of the BA534D display changes A programmable divider enables the puls...

Page 24: ...ng digit may be changed by pressing the Up or Down buttons When this digit is set as required pressing P will transfer control to the next digit When the flow rate corresponding to an output current o...

Page 25: ...mounting kits are available for securing the BA534D to a horizontal or vertical pipe BA392D Stainless steel bracket secured by two worm drive hose clips for 60 to 80mm outside diameter pipes BA393 He...

Page 26: ...ng commissioning 7 1 after commissioning 7 2 FILtEr 5 10 Flow indicator 2 3 Grand total 2 2 clear 5 14 Guarantee 7 5 HI LO 8 2 2 8 2 6 HStr 8 2 2 8 2 8 InPut 5 5 Installation 4 Instrument identificati...

Reviews: