6.1 Calibration procedure

The BA534D rate totaliser may be calibrated on-
site without disconnection from the power supply
or from the flowmeter.

Step 1

Enter the programming mode

Put the BA534D in the programming
mode by simultaneously pressing

  P

and

 E.  

Assuming a security code has

not already been entered the
instrument will respond by displaying
'UPdAtE' which is the first function in
the main menu.  See Fig 6.

Step 2

Select the interval between display
updates

With 'UPdAtE' displayed, press 

P

    to

reveal the existing interval between
display updates.  If this is not as
required, press the 

Up

 or 

Down

 button

until ‘0.5’ is displayed. (0.5 seconds i.e.
2 display updates per second).  Enter
the revised time and return to the
'UPdAtE' prompt in the main menu by
pressing 

E

.

Step 3

Select the type of input

Using the 

Up

 or 

Down

 button scroll

through the main menu until ‘InPut’ is
displayed, then press 

P

 to reveal the

existing setting.  Select  ‘COIL’, the
input for a magnetic pick-off, using the

Up

 or 

Down

 button and return to the

'InPut' prompt in the main menu by
pressing 

E

.

Note:

 The BA534D has two pairs of

input terminals. 3 & 4 for voltage inputs
and 5 & 6 for switch contact, proximity
detector or open collector inputs.
Ensure that the magnetic pick-off,
which has a voltage output, is
connected to terminals 3 & 4.

Step 4

Position rate & total decimal points

Select ‘d.P.’ from the main menu and
press 

P

.  The rate and total displays

will be activated with one digit of the
total display flashing.  Press the 

Up

 or

Down

 push-button until the third least

significant digit of the total display and
following decimal point are flashing.
This gives the required total display
resolution of 0.01

Press 

P

 to move control to the rate

display. Using the 

Up

 or 

Down

 push-

button move the flashing digit to the
least significant position (right hand
side) which will result in no decimal
point being displayed.

Finally press 

E

 to return to the 'd.P’.

prompt in the main menu.

Step 5

Enter the rate scaling factor

Select ‘SCALE-r’ from the main menu
and press 

P

 to show the current figure.

The K factor of the flowmeter in this
example is 1050.0* pulses per litre
which should be entered as the rate
scaling factor.

Firstly to position the decimal point,
operate the 

P

 push-button to move the

flashing digit to the second least
significant position.  Pressing the 

Up

and 

Down

 buttons simultaneously will

then position the decimal point in front
of the least significant digit.

Using the 

Up

 and 

Down

 buttons to

adjust each digit in turn and the 

P

button to transfer control between
digits, enter 1050.0     Finally return to
the 'SCALE-r' prompt in the main menu
by pressing 

E

.

Could also be entered as 1050

Step 6

Enter the rate timebase

Select ‘t-bASE’ from the main menu
and press 

P

 to show the current

setting.  In this example the rate
display is required in litres per hour.
Using the 

Up

 or 

Down

 button scroll

through the three options and select
‘tb-3600’ which will multiply the rate
display by 3600. Return to the ‘t-bASE’
prompt in the main menu by pressing

E

.

Step 7

Adjust the rate filter

The rate display filter parameters
should be adjusted experimentally after
installation to provide a stable rate
display with an acceptable step
response.

Summary of Contents for BA534D

Page 1: ...Issue 10 18th September 2008 BA534D Externally powered pulse input field mounting rate totaliser issue 10...

Page 2: ...Calibration procedure 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenance 7 5 Guarantee 7 6 Customer comments 8 Accessorie...

Page 3: ...ys further extend the applications Fig 1 Simplified block diagram of BA534D 2 1 Initalisation Each time power is applied to the instrument initalisation is performed After a short delay the following...

Page 4: ...et Activated when the total Annunciator display is being reset 3 SYSTEM DESIGN 3 1 Flow transmitter loops The BA534D may be connected to almost any pulse output flowmeter and calibrated to display rat...

Page 5: ...4D rate totaliser is housed in a robust IP66 glass reinforced polyester GRP enclosure incorporating an armoured glass window and stainless steel fittings It is suitable for exterior mounting in most i...

Page 6: ...rective 2004 108 EEC For specified immunity all wiring should be in screened twisted pairs and the BA534D earth terminal should be locally earthed or connected to an earthed cable screen To prevent ci...

Page 7: ...total scale factor SCALE t Total flow is continuously added to the grand total The BA534D displays real decimal points Moving the position of a decimal point in a scale factor will therefore affect th...

Page 8: ...8...

Page 9: ...2 Lower display Turns the lower display which normally shows rate on or off See section 5 7 Display Description of function SCALE r Rate Scale Factor Defines the arithmetic relationship between the p...

Page 10: ...s from a wide variety of sensors To define an input type select InPut from the menu and press P which will reveal the current input type Pressing Up or Down will scroll through the five options UOLts...

Page 11: ...igit can be changed by pressing the Up or Down buttons When this digit is correct pressing P will transfer control to the next digit To position the decimal point in the rate scale factor move the fla...

Page 12: ...r control to the next digit To position the decimal point in the scaling factor move the flashing digit to the left hand side of the required decimal point position and simultaneously press the Up and...

Page 13: ...git of the code can be changed using the Up and Down push buttons and the P button to move to the next digit When the required code has been entered press E to return to the main menu The revised secu...

Page 14: ...ush button until the third least significant digit of the total display and following decimal point are flashing This gives the required total display resolution of 0 01 Press P to move control to the...

Page 15: ...enter 1050000 Finally return to the SCALE t prompt in the main menu by pressing E Step 9 Define clip off In this example totalisation is to be inhibited at flow rates below 300 litres per hour Select...

Page 16: ...ode 1209 digit by digit The P button will transfer control between digits When the new code has been entered press E to return to the main programme menu Step 13 Return to the display mode Following c...

Page 17: ...20mA output calibration Alarms do not function Alarms have been disabled following calibration change Re enable both alarms 7 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING M...

Page 18: ...utputs WARNING These alarm outputs should not be used for critical safety applications such as a shut down system When the BA534D power supply is turned off or disconnected both BA534D alarm outputs w...

Page 19: ...19 Fig 10 Alarm menu structure...

Page 20: ...available on rate alarms See section 8 2 8 Summary of programmable alarm functions dELA Alarm delay time Adjusts the delay between the display equalling the setpoint and the alarm output being activa...

Page 21: ...st the hysteresis select HStr from the alarm menu and press P which will reveal the existing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next d...

Page 22: ...o the setpoints in the display mode by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with the security code set to 0000 8 2 12 Adjusting alarm setpoints from th...

Page 23: ...pulse output The output is a polarised current sink that closes for a programmable duration each time the least significant digit of the BA534D display changes A programmable divider enables the puls...

Page 24: ...ng digit may be changed by pressing the Up or Down buttons When this digit is set as required pressing P will transfer control to the next digit When the flow rate corresponding to an output current o...

Page 25: ...mounting kits are available for securing the BA534D to a horizontal or vertical pipe BA392D Stainless steel bracket secured by two worm drive hose clips for 60 to 80mm outside diameter pipes BA393 He...

Page 26: ...ng commissioning 7 1 after commissioning 7 2 FILtEr 5 10 Flow indicator 2 3 Grand total 2 2 clear 5 14 Guarantee 7 5 HI LO 8 2 2 8 2 6 HStr 8 2 2 8 2 8 InPut 5 5 Installation 4 Instrument identificati...

Reviews: