GX-TB-004EE
1-28
SX4750A Swinging-Bucket Rotor
Care and Maintenance
While Beckman Coulter has tested these methods and found that they do not damage the rotor or
components, no guarantee of decontamination is expressed or implied. Consult your laboratory
safety officer regarding the proper decontamination methods to use.
If the rotor or accessories are contaminated with toxic or pathogenic solutions, follow appropriate
decontamination procedures. Check
Chemical Resistances
to be sure the decontamination method
will not damage any part of the rotor.
Sterilization and Disinfection
•
The rotor yoke, buckets, adapters, and multiwell-plate carriers can be autoclaved at 121°C for
up to an hour. Blood bag cups, bucket and carrier covers, and Aerosolve cannisters can be
autoclaved at that temperature for about 15 minutes. (Do not autoclave Aerosolve cannisters
purchased prior to April 1993, or any components made of Noryl.) Remove the covers from the
buckets and place the rotor yoke, buckets, covers, and/or multiwell-plate carriers in the
autoclave upside down. Before autoclaving, remove the air-vent filter from each cover by
gently pushing it out from underneath the cover with a pencil or other non-metal tool that
will not scratch the cover material. After autoclaving, insert a new air-vent filter (368148) into
each cover.
•
Ethanol (70%)
*
or hydrogen peroxide (6%) may be used on all rotor components, including
those made of plastic. Bleach (sodium hypochlorite) may be used, but may cause discoloration
of anodized surfaces. Use the minimum immersion time for each solution, per laboratory
standards.
While Beckman Coulter has tested these methods and found that they do not damage the rotor or
components, no guarantee of sterility or disinfection is expressed or implied. When sterilization or
disinfection is a concern, consult your laboratory safety officer regarding proper methods to use.
Refer to publication IN-192, included in each box of tubes or bottles, for tube sterilization and
disinfection procedures.
Tube Breakage
CAUTION
To reduce the potential for corrosion, clean buckets or carriers thoroughly
immediately following a tube or well plate breakage.
If a glass tube breaks, remove the glass very carefully from the adapter and bucket. Imbedded glass
particles that remain in the bucket or adapters can cause tube failure during subsequent runs.
* In U.S., contact Nuclear Associates (New York); in Eastern Europe and Commonwealth States, contact Victoreen GmbH
(Munich); in South Pacific, contact Gammasonics Pty. Ltd. (Australia); in Japan, contact Toyo Medic Co. Ltd. (Tokyo).
† In U.S., contact Biodex Medical Systems (Shirley, New York); internationally, contact the U.S. office to find the dealer
closest to you.
* Flammability hazard. Do not use in or near operating ultracentrifuges.