GX-TB-004EE
1-12
SX4750A Swinging-Bucket Rotor
Loading the Buckets
Using Aerosolve Cannisters
CAUTION
Do not run chloroformed samples in Aerosolve cannisters. Chloroform vapors
can damage the cannister material.
Aerosolve cannisters can be used as wide-mouth bottles or with tubes in racks that are specially
designed to fit in the cannisters.
Table 1
lists the Aerosolve tube racks and the number and sizes of
tubes they accommodate. When Aerosolve cannisters are used in the SX4750A rotor buckets, pads
(361269) must be placed beneath the cannisters in the buckets to provide support and prevent the
cannisters from being damaged during centrifugation. These pads are included in the Aerosolve
cannister kits (359232). Bucket covers cannot be used with Aerosolve cannisters.
Using Blood Bag Cups
NOTE
This rotor is suitable for obtaining cell-free plasma or for cell packing.
The yellow 90-mm cup (356856) accommodates up to two single bags or one double pack; the orange
97-mm cup (356857) supports either one triple or one quad pack.
Do not pour liquid directly into blood
bag cups
. Fit blood bags into cups before loading the cups into the rotor buckets. Load the blood bag
cups as follows:
1
Load the bags as far down into the cups as possible.
NOTE
Make sure the bags stay as vertical as possible, with no folds at the top or corners. If folds are
present, blood cells could remain in the folds and then mix with the plasma when the bag is removed.
2
Sandwich the tubing between the blood bag and any transfer packs (see
Figure 3
).
1.