GX-TB-004EE
1-13
SX4750A Swinging-Bucket Rotor
Loading the Buckets
Figure 3
Blood Bag Loading Procedures
3
Make sure the loaded blood bag cups opposite each other on the rotor yoke are approximately
the same weight.
•
Soft, pliable balancing pads that will not puncture blood bags (see the
Supply List
) can be
used to maintain weight balance, if necessary.
4
Place loaded cups into rotor buckets.
•
To reduce the possibility of bag breakage, align the blood bag seam with the rotor pivot pins
(see
Figure 3
) with the label facing out.
5
Place bucket covers on rotor buckets to prevent blood bags or tubing from contacting the rotor
yoke during centrifugation.
1.
Blood Bag Label
2.
Blood Bag Seam Aligned with
Pivotal Axis of Bucket
3.
Ports
4.
Blood Bag Seam
5.
Blood Bag Cup
6.
Sandwich Coiled Tubing between
Transfer Bags and Filled Blood Bag