Description
Description
This Beckman Coulter rotor has been manufactured in an ISO 9001 or 13485 facility for use with the
specified Beckman Coulter centrifuges
.
The SX4750A is a four-place swinging bucket rotor. The rotor buckets carry adapters that allow
centrifugation of a wide range of tube and bottle sizes (from 1.5 mL reaction vials to 750-mL bottles)
or single- to quad-pack blood bags as well as multiwell plates in carriers. Special adapters are also
available that allow centrifugation of Corning canted-neck 75-cm
2
and 25-cm
2
cell culture flasks.
The rotor develops centrifugal forces that are suitable for rapidly sedimenting protein precipitates,
large particles, cells, and cell debris.
The rotor yoke is made of stainless steel. Black-anodized aluminum buckets and carriers can be run
by placing them over pivot pins on the arms of the yoke; they swing out to horizontal position
during centrifugation. Transparent covers made of high-impact plastic have been tested
*
to
demonstrate containment of microbiological aerosols under normal conditions of the associated
Beckman Coulter centrifuge when used and maintained as instructed. A captive tie-down device is
used to secure the rotor to the drive shaft during centrifugation.
The ARIES (Automated Rotor Imbalance Equilibrating System) technology provides imbalance
compensation for rotors with buckets that are up to 50 grams unbalanced.
The centrifuge identifies the rotor during the run by means of a magnetic sensor system in the
centrifuge rotor chamber and magnets imbedded in the rotor. The overspeed system ensures that
the rotor does not exceed its permitted speed.
Refer to the Warranty at the back of this manual for warranty information.
1.
Yoke
2.
Bucket
3.
Bottle
4.
Carrier
5.
Carriage
* Validation of microbiological containment was done at an independent third-party testing facility (Public Health
England). Improper use or maintenance may affect seal integrity and thus containment.