SX4750A Swinging-Bucket Rotor
Loading the Buckets
CAUTION
Buckets are designed to be used only with modular disk adapters, bottle adapters,
cell culture flask adapters, Aerosolve cannisters, and blood bag cups. Do not pour
samples directly into buckets, bottle sleeves, or blood bag cups. Do not load
bottles, cell culture flasks, or tubes directly into the buckets. Do not use labware
that is not specified for use in this rotor.
You can load buckets before or after they are installed on the rotor yoke. In either case, we
recommend filling the appropriate labware first and then loading the labware into the buckets.
Using Tubes
Assembling Modular Disk Adapters
Refer to
Table 1
to determine which color adapter is required for the tubes you are using and the
number of disks needed to support the length of the tubes.
1
Select the appropriate adapter base (from
Table 1
) and attach a bail to it (see
Figure 2
).
Figure 2
Assembling a Modular Disk Adapter
2
Position one of the disks so that its grooves are aligned with the bail.
•
Push the disk down until the bail snaps into the grooves.
1.
Bails
2.
Disks
3.
Base
1
3
2