Current Generation
MC4 User Guide
57
Warnings!
Make sure that you don't exceed the maximum current allowed
by the instrument under test.
If you open the current generation loop, MC4 tries to maintain
the current by increasing the output voltage. If you then close
the loop again, the current is first too high, but returns quickly
to the correct level. If this current peak could damage the
components of the loop, make sure that the loop will not open
or protect it against overcurrent. For the same reason, always
enter 0 mA output before connecting the loop.
Next
Current Measurement
on page 41.
Tools Menu
on page 67.
Stepping
on page 73
Ramping
on page 76.
Calibration
, see Part D.