Maintaining MC4's Instrument Database
172
MC4 User Guide
General Data Page
This is additional data typically received from a calibration software.
You also may enter/edit the data in MC4.
FIELD
DESCRIPTION
Work Order
The work order number/code for the cali-
bration of the current instrument.
Calibration Interval
The time interval between calibrations.
Number of
Calibration
Repeats,
- As Found Results
- As Left Results
Defines how many As Found and As Left
repeats should be done.
If a value other that zero is entered, after a
calibration is saved a window opens telling
you how many repeats are done and how
many should be done.
If the values are zeros, the above men-
tioned functionality is disabled.
Before Calibration
Note
Instructions helping you to prepare calibra-
tion. Free text field that expands with the
amount of text.
If this field contains text, it is shown after
the
F3
/
Start
button is pressed for the first
time.
See also chapter
Guidance Texts
on page
180.
Adjustment Note
Instructions helping you to adjust the in-
strument. Free text field that expands with
the amount of text.
If this field contains text, it is shown when
the adjustment window is opened
(
F1
/
Menu
,
Adjust Instrument
).
After Calibration
Note
Instructions helping you to finish the cali-
bration. Free text field that expands with
the amount of text.
If this field contains text, it is shown when
you close the calibration window.
Summary of Contents for MC4