Editing Instrument Data
MC4 User Guide
167
FIELD
(cont.)
DESCRIPTION
Supervised
Advancement to
Next Calibration
Point
If checked, advancing to next calibration
point requires user intervention.
Limited Checking
Range for Pass/Fail
Detection
If checked, you may enter a pass/fail de-
tection range that differs from the instru-
ment's input/output range. See also next
field.
Note that this setting is currently not
supported by calibration software. Thus
the pass/fail result may differ between
MC4 and the calibration software.
Checking Range,
0 % 100 %
Active only if previous Check box is set.
Enter the checking range for limited
pass/fail detection.
Note that the checking range is current-
ly not supported by calibration soft-
ware.
Procedure fields for switches
FIELD
DESCRIPTION
Automatic
Calibration Point
Acceptance
Check if you want MC4 to automatically
accept calibration points. See also chapter
Accepting Calibration Points Automati-
cally
on page 178.
Switch Ramp/Scan
Ranges
,
Set Point
0 %
,
100 %
Switch's set point should be within this
range. For best results, enter a range that
is symmetric in regard to the set point.
Continues on next page+